Schierbauer, JR; Günther, S; Haupt, S; Zimmer, RT; Zunner, BEM; Zimmermann, P; Wachsmuth, NB; Eckstein, ML; Aberer, F; Sourij, H; Moser, O.
Accuracy of Real Time Continuous Glucose Monitoring during Different Liquid Solution Challenges in Healthy Adults: A Randomized Controlled Cross-Over Trial.
Sensors (Basel). 2022; 22(9): 3104 Doi: 10.3390/s22093104 [OPEN ACCESS]
Web of Science PubMed PUBMED Central FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Moser Othmar
Co-Autor*innen der Med Uni Graz
Aberer Felix
Sourij Harald
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
Continuous glucose monitoring (CGM) represents an integral of modern diabetes management, however, there is still a lack of sensor performance data when rapidly consuming different liquids and thus changing total body water. 18 healthy adults (ten females, age: 23.1 ± 1.8 years, BMI 22.2 ± 2.1 kg·m-2) performed four trial visits consisting of oral ingestion (12 mL per kg body mass) of either a 0.9% sodium chloride, 5% glucose or Ringer's solution and a control visit, in which no liquid was administered (control). Sensor glucose levels (Dexcom G6, Dexcom Inc., San Diego, CA, USA) were obtained at rest and in 10-min intervals for a period of 120 min after solution consumption and compared against reference capillary blood glucose measurements. The overall MedARD [IQR] was 7.1% [3.3-10.8]; during control 5.9% [2.7-10.8], sodium chloride 5.0% [2.7-10.2], 5% glucose 11.0% [5.3-21.6] and Ringer's 7.5% [3.1-13.2] (p < 0.0001). The overall bias [95% LoA] was 4.3 mg·dL-1 [-19 to 28]; during control 3.9 mg·dL-1 [-11 to 18], sodium chloride 4.8 mg·dL-1 [-9 to 19], 5% glucose 3.6 mg·dL-1 [-33 to 41] and Ringer's solution 4.9 mg·dL-1 [-13 to 23]. The Dexcom G6 CGM system detects glucose with very good accuracy during liquid solution challenges in normoglycemic individuals, however, our data suggest that in people without diabetes, sensor performance is influenced by different solutions.

Find related publications in this database (Keywords)
continuous glucose monitoring
sensor accuracy
sodium chloride
glucose
Ringer's solution
healthy individuals
liquid consumption
© Med Uni Graz Impressum