Klarhöfer, M; Csapo, B; Balassy, C; Szeles, JC; Moser, E.
High-resolution blood flow velocity measurements in the human finger.
Magn Reson Med. 2001; 45(4):716-719
Doi: 10.1002/mrm.1096
Web of Science
PubMed
FullText
FullText_MUG
- Co-Autor*innen der Med Uni Graz
-
Csapo Bence Daniel
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
- MR phase contrast blood flow velocity measurements in the human index finger were performed with triggered, nontriggered, and cine acquisition schemes. A strong (G(max) = 200 mT/m), small bore (inner diameter 12 cm) gradient system inserted in a whole body 3 Tesla MR scanner allowed high-resolution imaging at short echo times, which decreases partial volume effects and flow artifacts. Arterial blood flow velocities ranging from 4.9-19 cm/sec were measured, while venous blood flow was significantly slower at 1.5-7.1 cm/sec. Taking into account the corresponding vessel diameters ranging from 800 microm to 1.8 mm, blood flow rates of 3.0-26 ml/min in arteries and 1.2-4.8 ml/min in veins are obtained. The results were compared to ultrasound measurements, resulting in comparable blood flow velocities in the same subjects. Magn Reson Med 45:716-719, 2001.
- Find related publications in this database (using NLM MeSH Indexing)
-
Blood Flow Velocity - physiology
-
Fingers - blood supply
-
Humans -
-
Magnetic Resonance Imaging - methods
-
Ultrasonography -
- Find related publications in this database (Keywords)
-
phase contrast
-
high resolution
-
blood flow velocity
-
human finger
-
3 Tesla