Gewählte Publikation:
Zobel, G; Ring, E; Trop, M.
Continuous arteriovenous renal replacement therapy.
Int J Artif Organs. 1987; 10(4):241-244
Doi: 10.1177/039139888701000407
Web of Science
PubMed
FullText
FullText_MUG
Google Scholar
- Co-Autor*innen der Med Uni Graz
-
Ring Ekkehard
-
Trop Marija
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
- As urea and creatinine clearances achieved by continuous arteriovenous hemofiltration tend to be low, efforts were made to increase urea and creatinine elimination by modifying the original continuous arteriovenous hemofiltration system. We investigated the urea and creatinine clearances of different continuous renal replacement therapy systems. Urea clearance achieved by spontaneous arteriovenous hemofiltration was 9.1 ml/min, suction-supported arteriovenous hemofiltration 15.3 ml/min, arteriovenous hemodiafiltration 15.4 to 19.3 ml/min, arteriovenous hemodialysis 18.9 to 22.1 ml/min. Creatinine clearances ranged between 11.1 (spontaneous arteriovenous hemofiltration) and 25.1 ml/min (arteriovenous hemodialysis). Continuous arteriovenous hemodialysis seems at present to be the ideal renal replacement therapy system to compensate azotemia in hypercatabolic patients. Its urea and creatinine clearances are high, its clinical tolerance good and it is simple and safe.
- Find related publications in this database (using NLM MeSH Indexing)
-
Child -
-
Creatinine - metabolism
-
Female - metabolism
-
Hemofiltration - metabolism
-
Humans - metabolism
-
Renal Dialysis - metabolism
-
Urea - metabolism