Pichler, W; Grechenig, W; Clement, H; Windisch, G; Tesch, NP.
Perforation of the third extensor compartment by the drill bit during palmar plating of the distal radius.
J Hand Surg Eur Vol. 2009; 34(3): 333-335.
Doi: 10.1177/1753193408099821
Web of Science
PubMed
FullText
FullText_MUG
- Leading authors Med Uni Graz
-
Pichler Wolfgang
- Co-authors Med Uni Graz
-
Clement Hans Gunther
-
Grechenig Wolfgang
-
Tesch Norbert
-
Windisch Gunther
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
- The frequency of penetration into the third extensor compartment when drilling the distal screw holes was assessed for four different palmar distal radius plates in 160 cadaver forearms. Penetration into this compartment occurred in 43%. Different plates had different penetration rates: 3.5 LCP four-hole locking T-plate: 20%, 3.5 LCP three-hole locking plate: 42.5%, 2.4 palmar LCP standard plate and the 2.4 palmar LCP buttress plate: 55%. When using a palmar plate on the distal radius, the surgeon risks penetrating into the third extensor compartment.
- Find related publications in this database (using NLM MeSH Indexing)
-
Aged -
-
Aged, 80 and over -
-
Bone Plates -
-
Cadaver -
-
Female -
-
Forearm -
-
Fracture Fixation, Internal - adverse effects Fracture Fixation, Internal - methods
-
Hand -
-
Hand Injuries - etiology
-
Humans -
-
Intraoperative Complications -
-
Male -
-
Middle Aged -
-
Radius Fractures - surgery
-
Surgical Instruments -
- Find related publications in this database (Keywords)
-
radius fracture
-
drilling
-
extensor pollicis rupture
-
palmar plating