Evaluation of the Gen-Probe PACE 2 and the Microtrak enzyme immunoassay for diagnosis of Chlamydia trachomatis in urogenital samples.
Sex Transm Dis. 1994; 21(1):26-30
Doi: 10.1097/00007435-199401000-00006
Web of Science
PubMed
FullText
FullText_MUG
Google Scholar
- Co-authors Med Uni Graz
-
Storch Maria
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
- BACKGROUND AND OBJECTIVES: The aim of the study was to investigate the value of the Gen-Probe PACE 2 assay for routine diagnosis of Chlamydia trachomatis in genital specimens of symptomatic and asymptomatic men and women patients. STUDY DESIGN: Samples were collected from 90 men and 299 women patients and tested by using the Gen-Probe assay and the EIA MicroTrak. Discrepant results were further analyzed by immunofluorescence, a second run of the Gen-Probe assay, and a probe competition assay (PCA) to establish the number of true positive and negative outcomes based on the two tests used. RESULTS: The overall prevalence of C. trachomatis was 8.5% in all patients tested (women: 3.7%, men: 13.3%) with an overall agreement of 95.4% between the two diagnostic methods. Of the 18 discordant results, 12 (67%) were considered to be false positive in the Gen-Probe assay and 3 (16%) false positive in the EIA. Two (11%) positive results were missed in the Gen-Probe assay and 1 (6%) in the EIA, all observed in female specimens. The sensitivities and specificities of the EIA were 91.7% and 100% for men and 100% and 99% for women, and for the Gen-Probe assay were 83.3% and 100% for men and 100% and 95.8% for women, respectively, when compared with true positive and true negative results. Although the predictive value for all positive results (PVP) was 88% for the EIA and 78.2% for the Gen-Probe assay, it was only 47.8% for positive female samples when using the Gen-Probe assay. CONCLUSION: The Gen-Probe assay revealed a sensitivity comparable with the EIA. The accuracy of test results provided by a single Gen-Probe assay was considerably lower than by Micro-Trak reducing the utility of PACE 2 as a diagnostic technique for Chlamydia diagnosis. Due to the high rate of false-positive samples in the Gen-Probe assay, positive results with a low value of relative light units have to be further analyzed by confirmation procedures.
- Find related publications in this database (using NLM MeSH Indexing)
-
Chlamydia Infections - diagnosis
-
Chlamydia trachomatis - diagnosis
-
Evaluation Studies as Topic - diagnosis
-
False Positive Reactions - diagnosis
-
Female - diagnosis
-
Humans - diagnosis
-
Immunoenzyme Techniques - diagnosis
-
Male - diagnosis
-
Sensitivity and Specificity - diagnosis