Sack, S; Heinzel, F; Dagres, N; Enger, S; Auricchio, A; Stellbrink, C; Neuzner, J; Potty, P; Maarse, A; Tockman, B; Michel, U; Erbel, R.
Stimulation of the left ventricle through the coronary sinus with a newly developed 'over the wire' lead system--early experiences with lead handling and positioning.
Europace. 2001; 3(4):317-323 Doi: 10.1053/eupc.2001.0185 [OPEN ACCESS]
Web of Science PubMed FullText FullText_MUG

 

Co-authors Med Uni Graz
Heinzel Frank
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
AIMS: This report describes the initial clinical results with a newly designed guiding catheter and an 'over the wire' pacing lead based on angiolasty technology to stimulate the left ventricle using the transvenous route via the coronary sinus (OTW-CV lead). METHODS AND RESULTS: In 75% of the 15 patients (6 males, 9 females, mean age of 53 +/- 9 years) with congestive heart failure, access to coronary sinus required less than 2 min; in one patient. the attempt failed. Mean R wave amplitudes plus or minus the standard deviation, measured at apical, mid-ventricular and basal positions the anterior (11.4 +/- 9.2, 10.8 +/- 6.2, 9.3 +/- 6.3 mV) and lateral or posterior veins (10.1 +/- 10.7, 8.6 +/- 6.4, 7.7 +/- 4.3 mV) showed a trend favouring the apex without statistical significance. Pacing impedance, measured at the same sites and vein tributaries, ranged from 670 +/- 191 to 915 +/- 145 ohms. Pacing thresholds measured at apical and mid ventricular sites were significantly lower than at the base in the anterior vein 2.5 +/- 2.8 and 2.8 +/- 1.8 vs 5.6 +/- 2.7 V at 0.5 ms, P<0.001). Thresholds in the lateral/posterior veins showed a similar trend but did not reach statistical significance (3.0 +/- 1.7, 3.6 +/- 1.4 +/- 1.8 V at 0.5 ms). In patients, in whom thresholds were determined in more than one vein, the 'best' mean threshold was 1.6 +/- 0.7 V. CONCLUSION: The new 'over the wire' lead and guiding catheter system allows uncomplicated access to the coronary sinus and the depth of the coronary vein tributaries. Left ventricular sensing and pacing thresholds are acceptable for chronic use in implanted cardiac rhythm management systems.
Find related publications in this database (using NLM MeSH Indexing)
Adult -
Cardiac Pacing, Artificial - methods
Coronary Vessels - anatomy and histology
Electrodes, Implanted - anatomy and histology
Equipment Design - anatomy and histology
Female - anatomy and histology
Heart Catheterization - anatomy and histology
Heart Failure, Congestive - therapy
Heart Ventricles - therapy
Humans - therapy
Male - therapy
Middle Aged - therapy
Pacemaker, Artificial - therapy

Find related publications in this database (Keywords)
left ventricular pacing
transvenous pacing
heart failure
coronary sinus