Medizinische Universität Graz Austria/Österreich - Forschungsportal - Medical University of Graz

Logo MUG-Forschungsportal

Gewählte Publikation:

SHR Neuro Krebs Kardio Lipid Stoffw Microb

Wackernagel, W; Ettinger, K; Weitgasser, U; Bakir, BG; Schmut, O; Goessler, W; Faschinger, C.
Opacification of a silicone intraocular lens caused by calcium deposits on the optic.
J Cataract Refract Surg. 2004; 30(2):517-520 Doi: 10.1016/S0886-3350(03)00672-2 (- Case Report)
Web of Science PubMed FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Wackernagel Werner
Co-Autor*innen der Med Uni Graz
Faschinger Christoph
Schmut Otto
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
We describe opacification of a plate-haptic silicone intraocular lens (IOL) caused by calcification in a diabetic patient with asteroid hyalosis. The IOL was explanted 48 months after uneventful phacoemulsification because opacification of the posterior surface was causing significant visual disturbance. Light and scanning electron microscopy and x-ray spectrometry of the explanted IOL showed the opacification consisted mainly of calcium and phosphate, presumably hydroxyapatite, in the form of precipitations on the posterior surface of the optic.
Find related publications in this database (using NLM MeSH Indexing)
Aged -
Biocompatible Materials - adverse effects
Calcinosis - etiology Calcinosis - pathology
Calcium - analysis
Device Removal -
Durapatite - analysis
Electron Probe Microanalysis -
Female -
Humans -
Lenses, Intraocular - adverse effects
Microscopy, Electron, Scanning -
Phosphates - analysis
Prosthesis Failure -
Silicone Elastomers - adverse effects