Gewählte Publikation:
Trathnigg, B; Kollroser, M.
Liquid chromatography of polyethers using universal detectors .5. Quantitative aspects in the analysis of low-molecular-mass poly(ethylene glycol)s and their derivatives by reversed-phase high-performance liquid chromatography with an evaporative light scattering detector
J CHROMATOGR A. 1997; 768(2): 223-238.
Doi: 10.1016/S0021-9673(97)00037-X
Web of Science
FullText
FullText_MUG
- Co-Autor*innen der Med Uni Graz
-
Kollroser Manfred
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
- The performance of an evaporative light scattering detector in liquid chromatography of poly(ethylene glycol)s and their mono- and dimethyl ethers is evaluated. The aspect of quantitation is critically examined for isocratic and gradient elution. Molar mass dependence of response factors for the different homologous series is studied at different evaporator temperatures, and compared to those from isocratic elution with density and refractive index detection.
- Find related publications in this database (Keywords)
-
evaporative light scattering detector
-
detection
-
LC
-
polyethers
-
poly(ethylene glycol)s