Nauck, MS; Nauck, MA; Nauck, MA; März, W; Wieland, H.
Six methods for the determination of C-peptide evaluated.
Clin Chem Lab Med. 1999; 37(7):745-751
Doi: 10.1515/CCLM.1999.115
Web of Science
PubMed
FullText
FullText_MUG
- Co-Autor*innen der Med Uni Graz
-
März Winfried
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
- Six assays for C-peptide were evaluated. Five assays were radioimmunoassays; one was a chemiluminescence assay. The within-run coefficient of variation ranged between 2.1 and 22.9%. The methods correlated well with each other (r = 0.925 to 0.978). There were, however, proportional biases between the assays, in particular in the higher concentration range. All assays were linear up to at least 4.5 ng/ml C-peptide concentration. All assays were easy to perform. However, the systematic bias between the assays demonstrates that further effort is needed to standardize the measurement of C-peptide. In the mean-time, we recommend that laboratories conducting C-peptide measurements establish local, method-specific reference ranges.
- Find related publications in this database (using NLM MeSH Indexing)
-
C-Peptide - blood
-
Chemiluminescent Measurements - blood
-
Female - blood
-
Humans - blood
-
Immunoassay - methods
-
Male - methods
-
Radioimmunoassay - methods