Gewählte Publikation:
Nicoletti, R; Vejda, M; Fueger, GF; Sorantin, E.
SPET quality control: a comprehensive test procedure for evaluating the rotation system parameters COR-offset, y-shift and detector tilt.
Nucl Med Commun. 1995; 16(2): 105-111.
Doi: 10.1097/00006231-199502000-00008
Web of Science
PubMed
FullText
FullText_MUG
- Führende Autor*innen der Med Uni Graz
-
Nicoletti Rudolf
- Co-Autor*innen der Med Uni Graz
-
Sorantin Erich
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
- This paper describes a method for SPET quality control that facilitates the determination of COR-offset, parallel y-shift and detector tilt by means of a single, comprehensive test using a special phantom consisting of three point sources. The evaluation of the method was performed by a series of measurements utilizing manually adjusted detector tilts and a simulation of y-shift by moving the phantom. The results show that y-shift and detector tilt are measured with reasonable accuracy, and that y-shift caused by variations in detector tilt can be differentiated from y-shift caused by other factors.
- Find related publications in this database (using NLM MeSH Indexing)
-
Evaluation Studies -
-
Humans -
-
Models, Structural -
-
Quality Control -
-
Rotation -
-
Tomography, Emission-Computed, Single-Photon - instrumentation