Medizinische Universität Graz - Research portal

Logo MUG Resarch Portal

Selected Publication:

SHR Neuro Cancer Cardio Lipid Metab Microb

Fortin, J; Marte, W; Grüllenberger, R; Hacker, A; Habenbacher, W; Heller, A; Wagner, Ch; Wach, P; Skrabal, F.
Continuous non-invasive blood pressure monitoring using concentrically interlocking control loops.
COMPUT BIOL MED. 2006; 36(9): 941-957. Doi: 10.1016/j.compbiomed.2005.04.003
Web of Science PubMed FullText FullText_MUG

 

Co-authors Med Uni Graz
Marte Wolfgang
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
A new method and apparatus for non-disruptive blood pressure (BP) recording in the finger based on the vascular unloading technique is introduced. The instrument, in contrast to intermittent set point readjustments of the conventional vascular unloading technique, delivers BP without interruptions, thus refining the Penáz' principle. The method is based on concentrically interlocking control loops for correct long-term tracing of finger BP, including automatic set point adaptation, light control and separate inlet and outlet valves for electro-pneumatic control. Examples of long-term BP recordings at rest and during autonomic function tests illustrate the potential of the new instrument.
Find related publications in this database (using NLM MeSH Indexing)
Blood Pressure Monitors -
Blood Volume - physiology
Carotid Sinus - physiology
Electronics, Medical - instrumentation
Equipment Design - instrumentation
Femoral Artery - physiology
Fingers - blood supply
Fuzzy Logic - blood supply
Humans - blood supply
Plethysmography - instrumentation
Pulsatile Flow - physiology
Radial Artery - physiology
Respiration - physiology
Transducers, Pressure - physiology
Transillumination - instrumentation
Valsalva Maneuver - physiology

Find related publications in this database (Keywords)
non-invasive finger blood pressure
vascular unloading technique
intra-arterial pressure
Penaz principle
© Med Uni GrazImprint