Fortin, J; Marte, W; Grüllenberger, R; Hacker, A; Habenbacher, W; Heller, A; Wagner, Ch; Wach, P; Skrabal, F.
Continuous non-invasive blood pressure monitoring using concentrically interlocking control loops.
COMPUT BIOL MED. 2006; 36(9): 941-957.
Doi: 10.1016/j.compbiomed.2005.04.003
Web of Science
PubMed
FullText
FullText_MUG
- Co-Autor*innen der Med Uni Graz
-
Marte Wolfgang
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
- A new method and apparatus for non-disruptive blood pressure (BP) recording in the finger based on the vascular unloading technique is introduced. The instrument, in contrast to intermittent set point readjustments of the conventional vascular unloading technique, delivers BP without interruptions, thus refining the Penáz' principle. The method is based on concentrically interlocking control loops for correct long-term tracing of finger BP, including automatic set point adaptation, light control and separate inlet and outlet valves for electro-pneumatic control. Examples of long-term BP recordings at rest and during autonomic function tests illustrate the potential of the new instrument.
- Find related publications in this database (using NLM MeSH Indexing)
-
Blood Pressure Monitors -
-
Blood Volume - physiology
-
Carotid Sinus - physiology
-
Electronics, Medical - instrumentation
-
Equipment Design - instrumentation
-
Femoral Artery - physiology
-
Fingers - blood supply
-
Fuzzy Logic - blood supply
-
Humans - blood supply
-
Plethysmography - instrumentation
-
Pulsatile Flow - physiology
-
Radial Artery - physiology
-
Respiration - physiology
-
Transducers, Pressure - physiology
-
Transillumination - instrumentation
-
Valsalva Maneuver - physiology
- Find related publications in this database (Keywords)
-
non-invasive finger blood pressure
-
vascular unloading technique
-
intra-arterial pressure
-