Medizinische Universität Graz Austria/Österreich - Forschungsportal - Medical University of Graz

Logo MUG-Forschungsportal

Gewählte Publikation:

Halwachs-Baumann, G; Katzensteiner, S; Schnedl, W; Pürstner, P; Pieber, T; Wilders-Truschnig, M.
Comparative evaluation of three assay systems for automated determination of hemoglobin A1c.
Clin Chem. 1997; 43(3):511-517 Doi: 10.1093/clinchem/43.3.511 [OPEN ACCESS]
Web of Science PubMed FullText FullText_MUG Google Scholar

 

Führende Autor*innen der Med Uni Graz
Baumann Gabriele
Co-Autor*innen der Med Uni Graz
Pieber Thomas
Pürstner Peter
Schnedl Wolfgang
Truschnig-Wilders Martini
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
We evaluated three newly introduced systems for automated determinations of hemoglobin (Hb) A1c, which allow the processing of large amounts of samples in a routine clinical laboratory. We compared these methods--the Variant HPLC, the Hi-Auto A1c analyzer system, and the Roche immunoassay--with the Diamat HPLC system. All showed good precision and good concordance with the Diamat HPLC. The reference range for Hb A1c has to be determined by the laboratory for each assay system. Interference study showed no statistically significant influence of anemia, polycythemia, rheumatoid factor, or chronic hemodialysis, although individual Hb A1c values can be influenced by polycythemia (when measured with the Hi-Auto A1c analyzer) and by chronic hemodialysis (when measured with the Variant HPLC). HPLC was not suitable for measuring Hb A1c in the examined cases of hemoglobin variants; assaying fructosamine seems to be better for monitoring these patients.
Find related publications in this database (using NLM MeSH Indexing)
Anemia - blood
Arthritis, Rheumatoid - blood
Autoanalysis - methods
Chromatography, High Pressure Liquid - methods
Diabetes Mellitus, Type 1 - blood
Diabetes Mellitus, Type 2 - blood
Glycated Hemoglobin A - analysis
Humans -
Immunoassay - methods
Kidney Failure, Chronic - blood
Kidney Failure, Chronic - therapy
Polycythemia - blood
Reference Values -
Renal Dialysis -
Reproducibility of Results -
Sensitivity and Specificity -

Find related publications in this database (Keywords)
Chromatography
Liquid
Immunoassay
Intermethod Comparison
Variation
Source of
Hemoglobin Variants
© Med Uni Graz Impressum