Assessment of the single-occlusion technique for measurements of respiratory mechanics and respiratory drive in healthy term neonates using a commercially available computerized pulmonary function testing system.
Biol Neonate. 2003; 83(2):117-122 Doi: 10.1159/000067957
Web of Science PubMed FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Reiterer Friedrich
Co-Autor*innen der Med Uni Graz
Müller Wilhelm
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
In this study, a critical assessment of the single-occlusion technique as a means of measuring passive respiratory mechanics and respiratory drive (P0.1) was performed in nonintubated spontaneously breathing healthy term neonates using commercially available computerized equipment (PEDS system). In general, we found that quality parameters only partially conformed to the international established standards for measuring passive respiratory mechanics. There was a failure rate of technically acceptable occlusions of about 50% for respiratory mechanics measurements and 20% for P0.1 measurements. Intersubject variability of the data was between 10 and 40%. After studying different lengths of occlusion times (Tocc; 0.2, 0.3, 0.4 and 0.5 s), it was found that acceptable occlusions for passive respiratory mechanics were obtained only if the expiratory Tocc exceeded 0.2 s. Increasing the Tocc had no effect on the absolute values of respiratory mechanics but showed a significant trend towards more technically acceptable occlusions. We speculate that a Tocc of at least 0.4 s may be a prerequisite for obtaining reliable results for respiratory mechanics in healthy term neonates.
Find related publications in this database (using NLM MeSH Indexing)
Automatic Data Processing -
Humans -
Infant, Newborn - physiology
Respiratory Function Tests - instrumentation
Respiratory Mechanics - instrumentation

Find related publications in this database (Keywords)
single-occlusion technique
respiratory mechanics
occlusion time
P0.1
neonates
© Med Uni Graz Impressum