Nepp, J; Krepler, K; Jandrasits, K; Hauff, W; Hanselmayer, G; Velikay-Parel, M; Ossoinig, KC; Wedrich, A.
Biometry and refractive outcome of eyes filled with silicone oil by standardized echography and partial coherence interferometry.
GRAEF ARCH CLIN EXP OPHTHAL 2005 243: 967-972.
Doi: 10.1007/s00417-004-1117-2
Web of Science
PubMed
FullText
FullText_MUG
- Co-authors Med Uni Graz
-
Velikay-Parel Michaela
-
Wedrich Andreas
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
- BACKGROUND: We evaluated the validity of calculations for refractive outcome in cataract surgery in silicone oil-filled eyes. The retrosilicone space (RSS) was included in these calculations. METHODS: In a prospective study the axial length (AL) of silicone oil-filled eyes was measured. with standardized A-scan echography (SAE) and partial coherence interferometry (PCI). Meldrum's formula was used to transform the velocity of ultrasound within the vitreous cavity. To investigate whether refractive outcome can be calculated accurately, we assessed the difference between precalculated and final refractive outcome. Furthermore, we determined the advantages and disadvantages of SAE and PCI. A minor aim was to assess whether the AL of the two eyes differed significantly. RESULTS: In 85% of 117 eyes the difference between precalculated and postsurgical refraction was smaller than 1 diopter spherical and statistically not significant (p>0.2). The mean AL was 24.1 mm (range 20.0-31.4 mm). The difference in outcome between the two methods was without statistical significance: the AL difference was 0.4 (+/-2.6) mm on measurement with SAE and 0.04 (+/-0.46) mm with PCI. PCI has the advantage that it can be performed more easily, without contact, while echography is advantageous in the presence of advanced cataracts. In supine position an oil-free fluid space behind the silicone oil was detected with echography. The mean dimension of this space was 1.9 (+/-0.67) mm and it was taken into consideration for IOL calculation. The mean AL difference between the two eyes was 0.4 mm, but the difference was greater than 1 mm in 26% of the patients. CONCLUSION: The AL of eyes filled with silicone oil can be measured reliably with SAE and PCI. In supine position the RSS has to be considered to obtain more accurate IOL calculations.
- Find related publications in this database (using NLM MeSH Indexing)
-
Adolescent -
-
Adult -
-
Aged -
-
Aged, 80 and over -
-
Biometry - methods
-
Child - methods
-
Comparative Study - methods
-
Eye - ultrasonography
-
Female - ultrasonography
-
Follow-Up Studies - ultrasonography
-
Humans - ultrasonography
-
Injections - ultrasonography
-
Interferometry - methods
-
Lens Implantation, Intraocular - methods
-
Male - methods
-
Middle Aged - methods
-
Phacoemulsification - methods
-
Prospective Studies - methods
-
Refraction, Ocular - physiology
-
Silicone Oils - administration and dosage
-
Treatment Outcome - administration and dosage
-
Vitreous Body - administration and dosage
- Find related publications in this database (Keywords)
-
intravitreal silicone oil
-
biometry
-
standardized echography
-
partial coherence interferometry
-
retrosilicone space