Zweiker, R; Schumacher, M; Fruhwald, FM; Watzinger, N; Klein, W.
Comparison of wrist blood pressure measurement with conventional sphygmomanometry at a cardiology outpatient clinic.
J Hypertens. 2000; 18(8):1013-1018 Doi: 10.1097%2F00004872-200018080-00004
Web of Science PubMed FullText FullText_MUG

 

Leading authors Med Uni Graz
Zweiker Robert
Co-authors Med Uni Graz
Fruhwald Friedrich
Schumacher Martin
Watzinger Norbert
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
OBJECTIVE: Oscillometric measurement of blood pressure at the wrist is becoming a widely used method for detection of hypertension and its control by treatment. The objective of the present study was to evaluate accuracy and suitability of wrist measurement in a clinical routine setting. PATIENTS AND METHODS: A series of 333 consecutive patients admitted to our cardiology outpatient clinic were included. Blood pressure was measured at both upper arms according to World Health Organization-International Society of Hypertension guidelines. Oscillometric measurement was performed at the contralateral wrist simultaneously. Blood pressure readings were taken by an oscillometric device applied at the wrist ('Klock'; Industrielle Entwicklung Medizintechnik, Stolberg, Germany) and a conventional mercury sphygmomanometer applied at the upper arm. RESULTS: Seventy-eight patients were excluded due to differences in blood pressure > 5 mmHg between both upper arms or due to 'error' messages of the wrist device. The data of the remaining 255 patients (149 males; mean age, 65 +/- 13 years; range, 18-95 years) are presented. Mean conventional blood pressure was significantly lower compared with the wrist device (137 +/- 20/80 +/- 11 mmHg versus 153 +/- 28/87 +/- 18 mmHg; P < 0.001 and P < 0.001). The mean difference was 16 +/- 25/6 +/- 17 mmHg. In clinical terms, differences in blood pressure exceeding +/-20/+/-10 mmHg reflecting classification of hypertension are considered important. Measurements of 101 (40%) patients were within these limits. Systolic readings of 110 (43%) and diastolic readings of 117 patients (46%) were beyond this scope. CONCLUSION: Due to low reliability of wrist blood pressure measurement, it cannot compete with the upper arm standard procedure. If ever, it should only be used if test readings in an individual comparing wrist and upper arm measurement show differences within a range of +/-20/+/-10 mmHg.
Find related publications in this database (using NLM MeSH Indexing)
Adolescent -
Adult -
Aged -
Aged, 80 and over -
Blood Pressure - physiology
Blood Pressure Determination - instrumentation
Cardiology Service, Hospital - instrumentation
Female - instrumentation
Humans - instrumentation
Male - instrumentation
Middle Aged - instrumentation
Nurses - instrumentation
Outpatient Clinics, Hospital - instrumentation
Wrist - physiology

Find related publications in this database (Keywords)
blood pressure measurement
sphygmomanometer
oscillometry
wrist
clinical study