Medizinische Universität Graz - Research portal

Logo MUG Resarch Portal

Selected Publication:

Leis, HJ; Fauler, G; Raspotnig, G; Windischhofer, W.
Quantitative analysis of morphine in human plasma by gas chromatography-negative ion chemical ionization mass spectrometry.
J Chromatogr B Biomed Sci Appl. 2000; 744(1):113-119 Doi: 10.1016%2FS0378-4347%2800%2900237-1
Web of Science PubMed FullText FullText_MUG

 

Leading authors Med Uni Graz
Leis Hans-Joerg
Co-authors Med Uni Graz
Fauler Günter
Windischhofer Werner
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
A sensitive and specific method for the quantitative determination of morphine in human plasma is presented. Morphine was extracted from plasma by solvent extraction with ethyl acetate and derivatized to its heptafluorobutyrate derivative. The derivatives were measured by gas chromatography-negative ion chemical ionization mass spectrometry without any further purification. Using this detection mode, a diagnostic useful fragment ion at m/z 637 is obtained at high relative abundance. Deuterated morphine was used as an internal standard. Calibration graphs were linear within a range of 0.78 ng/ml and 50 ng/ml. Inter-assay precision was 2.3% (2.85 ng/ml) and 1.4% (14.25 ng/ml), intra-assay variability was found to be 1.5% (3.71 ng/ml) and 0.5% (14.54 ng/ml). Accuracy showed deviations of -9.3% (2.85 ng/ml) and -4.2% (14.25 ng/ml). The method is rugged and robust and has been applied to the batch analysis of morphine during pharmacokinetic profiling of the drug.
Find related publications in this database (using NLM MeSH Indexing)
Analgesics, Opioid - blood
Gas Chromatography-Mass Spectrometry - methods
Humans - methods
Morphine - blood
Reproducibility of Results - blood

Find related publications in this database (Keywords)
morphine
© Med Uni GrazImprint