Assessing the variability of 23 Na MRI in skeletal muscle tissue: Reproducibility and repeatability of tissue sodium concentration measurements in the lower leg at 3 T.
NMR Biomed. 2020; 33(5): e4279 Doi: 10.1002/nbm.4279
Web of Science PubMed FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Gerhalter Teresa
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
The goal of this study was to evaluate the reproducibility and repeatability of tissue sodium concentration (TSC) measurements using 23 Na MRI in skeletal muscle tissue. 23 Na MRI was performed at 3 T on the right lower leg of eight healthy volunteers (aged 28 ± 4 years). The examinations were repeated at the same site after ~ 22 weeks to assess the variability over a medium-term period. Additionally, they were scanned at a second site shortly before or shortly after the first visit (within 3 weeks) to evaluate the inter-site reproducibility. Moreover, we analysed the effect of B0 correction on the variability. Coefficients of variations (CVs) from mean TSC values as well as Bland-Altman plots were used to assess intra-site repeatability and inter-site reproducibility. In phantom measurements, the B0 correction improved the quantitative accuracy. We observed differences of up to 4.9 mmol/L between the first and second visit and a difference of up to 3.7 mmol/L between the two different sites. The CV for the medium-term repeatability was 15% and the reproducibility CV was 9%. The Bland-Altman plots indicated high agreement between the visits in all muscle regions. The systematic bias of -0.68 mmol/L between site X and Y (P = 0.03) was slightly reduced to -0.64 mmol/L after B0 correction (P = 0.04). This work shows that TSC measurements in healthy skeletal muscle tissue can be performed with good repeatability and reproducibility, which is of importance for future longitudinal or multicentre studies.
Find related publications in this database (using NLM MeSH Indexing)
Adult - administration & dosage
Humans - administration & dosage
Leg - physiology
Magnetic Resonance Imaging - administration & dosage
Male - administration & dosage
Muscle, Skeletal - diagnostic imaging
Reproducibility of Results - administration & dosage
Signal-To-Noise Ratio - administration & dosage
Sodium - analysis

Find related publications in this database (Keywords)
quantification
repeatability
reproducibility
skeletal muscle tissue
sodium MRI
total tissue sodium content