Medizinische Universität Graz Austria/Österreich - Forschungsportal - Medical University of Graz

Logo MUG-Forschungsportal

Gewählte Publikation:

SHR Neuro Krebs Kardio Lipid Stoffw Microb

He, X; Bender, M; Gross, C; Narayanaswamy, K; Laufer, G; Jakubek, S; Bonderman, D; Roehrich, M; Karner, B; Zimpfer, D; Granegger, M.
Left Atrial Decompression With the HeartMate3 in Heart Failure With Preserved Ejection Fraction: Virtual Fitting and Hemodynamic Analysis.
ASAIO J. 2024; 70(2):107-115 Doi: 10.1097/MAT.0000000000002074
Web of Science PubMed FullText FullText_MUG

 

Co-Autor*innen der Med Uni Graz
Laufer Günther
Zimpfer Daniel
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
Effective treatment of heart failure with preserved ejection fraction (HFpEF) remains an unmet medical need. Although left atrial decompression using mechanical circulatory support devices was previously suggested, the heterogeneous HFpEF population and the lack of tailored devices have prevented the translation into clinical practice. This study aimed to evaluate the feasibility of left atrial decompression in HFpEF patients with a HeartMate 3 (HM3, Abbott Inc, Chicago, USA) in silico and in vitro . Anatomic compatibility of the HM3 pump was assessed by virtual device implantation into the left atrium through the left atrial appendage (LAA) and left atrial posterior wall (LAPW) of 10 HFpEF patients. Further, the efficacy of left atrial decompression was investigated experimentally in a hybrid mock loop, replicating the hemodynamics of an HFpEF phenotype at rest and exercise conditions. Virtual implantation without substantial intersection with surrounding tissues was accomplished through the LAA in 90% and 100% through the LAPW. Hemodynamic analysis in resting conditions demonstrated normalization of left atrial pressures without backflow at a pump speed of around 5400 rpm, whereas a range of 6400-7400 rpm was required during exercise. Therefore, left atrial decompression with the HM3 may be feasible in terms of anatomic compatibility and hemodynamic efficacy.
Find related publications in this database (using NLM MeSH Indexing)
Humans - administration & dosage
Heart Failure - therapy
Atrial Appendage - administration & dosage
Stroke Volume - administration & dosage
Heart Atria - surgery
Hemodynamics - administration & dosage
Decompression - administration & dosage
Ventricular Function, Left - administration & dosage

Find related publications in this database (Keywords)
HFpEF
HeartMate3
virtual fitting
hemodynamics
hybrid mock loop
© Med Uni Graz Impressum