Anatomical Compliance of Cavopulmonary Assist Device Designs: A Virtual Fitting Study in Fontan Patients.
ASAIO J. 2023; 69(11):1016-1024 Doi: 10.1097/MAT.0000000000002013
Web of Science PubMed FullText FullText_MUG

 

Co-authors Med Uni Graz
Zimpfer Daniel
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
Several device designs for cavopulmonary mechanical circulatory support (MCS) are under investigation, however, challenged by the Fontan population's heterogeneity in size, cardiovascular and thoracic anatomy. This study aimed to preclinically assess the anatomical compliance of proposed device designs in silico. Representative double- and single-outlet cavopulmonary assist device (CPAD) designs were virtually implanted into CT imaging data of 10 patients previously palliated with total cavopulmonary connection (TCPC) for functionally univentricular hearts. Anatomical device compatibility was characterized concerning pump proximity to cardiovascular, respiratory and thoracic structures, as well as pump in- and outflow graft configuration. In 10 Fontan patients with a median age of 10.4 years (interquartile range [IQR] 5.0-15.3 years) and a median body surface area of 1.09 m2 (IQR 0.76-1.28 m2), implantation of a double-outlet CPAD was feasible in 1 patient (10%). In all other, adverse device intersection with the trachea and (neo-)aorta, or posterior pulmonary artery outflow graft kinking were observed. A single-outlet design permitted enhanced device mobilization adapting to individual anatomical conditions, resulting in device fit in nine of 10 patients (90%). Despite vast anatomical variations among single ventricle patients, a single-outlet device design may provide intracorporeal cavopulmonary MCS to a broad spectrum of failing Fontan patients.
Find related publications in this database (using NLM MeSH Indexing)
Humans - administration & dosage
Child, Preschool - administration & dosage
Child - administration & dosage
Adolescent - administration & dosage
Pulmonary Artery - administration & dosage
Aorta - administration & dosage
Body Surface Area - administration & dosage
Patient Compliance - administration & dosage
Patients - administration & dosage

Find related publications in this database (Keywords)
total cavopulmonary connection
Fontan circulation
Fontan failure
mechanical circulatory support
cavopulmonary assist device
virtual fitting