Medizinische Universität Graz - Research portal

Logo MUG Resarch Portal

Selected Publication:

SHR Neuro Cancer Cardio Lipid Metab Microb

Wendl, T; Wendl, B.
A comparative in vitro study of bond strength of adhesive precoated and not precoated brackets
INF ORTHOD KIEFEROR. 2023; 55(03): 157-160. Doi: 10.1055/a-2144-0960
Web of Science FullText FullText_MUG

 

Leading authors Med Uni Graz
Wendl Brigitte
Altmetrics:

Dimensions Citations:
Plum Analytics:


Scite (citation analytics):

Abstract:
In this study, the bond strength of adhesive precoated brackets was compared to brackets without adhesive coating using the conventional bonding method.Shear testing was performed with the universal mechanical testing machine at a feed rate of 0.5 mm/min and ARI Index was determined by stereo microscope (10-66 fold magnification). Statistical analysis was carried out using the SPSS program. The level of significance was established at p<0.05. The T-Test for independent random samples was employed when comparing two mean values.Both groups achieved the minimum bond strength of 5-8 MPa and there was only a significant difference between the lower incisors groups.ARI: In 20 of 30 cases in the bracket group without adhesive coating>90% of the composite remained on the enamel. The APC System also showed this fracture mode but a smaller tendency towards bracket-adhesion failure was recognizable.

Find related publications in this database (Keywords)
bond strength
adhesive precoated brackets
Adhesive Remnant Index [ARI]
© Med Uni GrazImprint