Restoring tibial obliquity for kinematic alignment in total knee arthroplasty: conventional versus patient-specific instrumentation.
Arch Orthop Trauma Surg. 2023; 143(9):5867-5872 Doi: 10.1007/s00402-023-04845-7 [OPEN ACCESS]
Web of Science PubMed PUBMED Central FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Sadoghi Patrick
Smolle Maria Anna
Co-Autor*innen der Med Uni Graz
Koutp Amir
Leithner Andreas
Leitner Lukas
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
INTRODUCTION: In total knee arthroplasty (TKA), tibial obliquity-restoration using kinematic alignment (KA) poses a major difference to conventional mechanical alignment. This study aimed at analysing the accuracy of conventional instrumentation (CI) versus patient-specific instrumentation (PSI) to restore anatomic tibial obliquity measured by the medial proximal tibial angle (MPTA) on conventional X-rays. MATERIALS AND METHODS: One-hundred patients were randomized to receive CI (n = 50) or PSI (n = 50) for TKA. Further 100 patients received CI without randomisation, resulting in 200 patients in total (127 women, mean age: 70.7 (range: 48-90 years). Pre- and postoperative X-rays were measured twice by two observers with a 2-week break in-between. Inter- and intraclass correlations were calculated and postoperative tibial obliquity compared to preoperative anatomy. RESULTS: In 150 patients with CI, no case with tibial obliquity-deviation greater than 2° was found, whilst 21.3% (n = 32) and 0.7% (n = 1) of cases and had a deviation of 0°-1°, and 1°-2°, respectively. In the remaining 78.0% (n = 117), tibial obliquity was restored. In 50 patients with PSI, no single case with a deviation greater than 1° was found. Sixty percent (n = 30) had a deviation of 0°-1°. In the remaining 40.0% (n = 20), no deviation from preoperative measurements was found. Consequently, CI resulted in a significantly smaller change in tibial obliquity from preoperative to postoperative than PSI (p < 0.001). Inter- and intra-class correlations showed a substantial agreement (any ICC > 0.90). CONCLUSION: Both conventional and patient-specific instrumentation revealed adequate results with respect to restoring tibial obliquity in kinematically aligned TKA, with conventional instrumentation achieving superior results.
Find related publications in this database (using NLM MeSH Indexing)
Humans - administration & dosage
Female - administration & dosage
Aged - administration & dosage
Arthroplasty, Replacement, Knee - methods
Biomechanical Phenomena - administration & dosage
Osteoarthritis, Knee - surgery
Tibia - surgery
Radiography - administration & dosage
Knee Joint - diagnostic imaging, surgery
Knee Prosthesis - administration & dosage

Find related publications in this database (Keywords)
Total knee arthroplasty
Kinematic alignment
Patient-specific instrumentation
Conventional instrumentation