Fleischhackl, R; Singer, F; Nitsche, W; Gamperl, G; Roessler, B; Arrich, J; Fleischhackl, S; Losert, H; Sterz, F; Mittlboeck, M; Hoerauf, K.
Influence of electromagnetic fields on function of automated external defibrillators.
Acad Emerg Med. 2006; 13(1):1-6 Doi: 10.1197/j.aem.2005.07.036
Web of Science PubMed FullText FullText_MUG

 

Co-Autor*innen der Med Uni Graz
Singer Florian
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
OBJECTIVES: In this study, the authors tested whether electromagnetic interference (EMI) is able to impair correct electrocardiogram analysis and produce false-positive shock advice from automated external defibrillators (AEDs) when the true rhythm is sinus. METHODS: Nineteen healthy subjects were used to test five AEDs available on the Austrian market in a prospective, open, and sequence-randomized study. The primary outcome variable was the absolute number of shocks advised in the presence of EMI. The secondary outcome was the number of impaired analyses caused by incorrectly detected patient movements or electrode failure. RESULTS: Of 760 tests run, 18 (2.37%) cases of false-positive results occurred, and two of five AEDs recommended shocks in the presence of sinus rhythm. Of 760 tests run, no electrode failures occurred. There were 27 occurrences (3.55%) of motion detected by an AED in the presence of strong electromagnetic fields. CONCLUSIONS: AED models differ in their response to EMI; it may be useful to consider specific safety requirements for areas with such fields present. Working personnel and emergency medical services staff should be informed about potential risks and the possible need for patient evacuation before AEDs are attached and shock recommendations are followed.
Find related publications in this database (using NLM MeSH Indexing)
Adult - administration & dosage
Electric Countershock - instrumentation
Electrocardiography - instrumentation
Electromagnetic Fields - administration & dosage
Equipment Failure Analysis - administration & dosage
False Positive Reactions - administration & dosage
Female - administration & dosage
Humans - administration & dosage
Male - administration & dosage
Movement - administration & dosage
Posture - administration & dosage
Probability - administration & dosage
Prospective Studies - administration & dosage
Reference Values - administration & dosage

Find related publications in this database (Keywords)
automated external defibrillator
electromagnetic interference
public access defibrillation
© Med Uni Graz Impressum