Singer, F; Houltz, B; Latzin, P; Robinson, P; Gustafsson, P.
A realistic validation study of a new nitrogen multiple-breath washout system.
PLoS One. 2012; 7(4):e36083 Doi: 10.1371/journal.pone.0036083 [OPEN ACCESS]
Web of Science PubMed PUBMED Central FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Singer Florian
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
BACKGROUND: For reliable assessment of ventilation inhomogeneity, multiple-breath washout (MBW) systems should be realistically validated. We describe a new lung model for in vitro validation under physiological conditions and the assessment of a new nitrogen (N(2))MBW system. METHODS: The N(2)MBW setup indirectly measures the N(2) fraction (F(N2)) from main-stream carbon dioxide (CO(2)) and side-stream oxygen (O(2)) signals: F(N2) = 1-F(O2)-F(CO2)-F(Argon). For in vitro N(2)MBW, a double chamber plastic lung model was filled with water, heated to 37°C, and ventilated at various lung volumes, respiratory rates, and F(CO2). In vivo N(2)MBW was undertaken in triplets on two occasions in 30 healthy adults. Primary N(2)MBW outcome was functional residual capacity (FRC). We assessed in vitro error (√[difference](2)) between measured and model FRC (100-4174 mL), and error between tests of in vivo FRC, lung clearance index (LCI), and normalized phase III slope indices (S(acin) and S(cond)). RESULTS: The model generated 145 FRCs under BTPS conditions and various breathing patterns. Mean (SD) error was 2.3 (1.7)%. In 500 to 4174 mL FRCs, 121 (98%) of FRCs were within 5%. In 100 to 400 mL FRCs, the error was better than 7%. In vivo FRC error between tests was 10.1 (8.2)%. LCI was the most reproducible ventilation inhomogeneity index. CONCLUSION: The lung model generates lung volumes under the conditions encountered during clinical MBW testing and enables realistic validation of MBW systems. The new N(2)MBW system reliably measures lung volumes and delivers reproducible LCI values.
Find related publications in this database (using NLM MeSH Indexing)
Adult - administration & dosage
Aged - administration & dosage
Breath Tests - methods
Female - administration & dosage
Functional Residual Capacity - physiology
Humans - administration & dosage
Lung - anatomy & histology, physiology
Male - administration & dosage
Middle Aged - administration & dosage
Models, Biological - administration & dosage
Nitrogen - analysis
Organ Size - administration & dosage
Reproducibility of Results - administration & dosage
Young Adult - administration & dosage

© Med Uni Graz Impressum