Selected Publication:
SHR
Neuro
Cancer
Cardio
Lipid
Metab
Microb
Weis, J; Geiger, R; Kilo, J; Zimpfer, D.
Cormatrix® for vessel reconstruction in paediatric cardiac surgery-a word of caution.
Interact Cardiovasc Thorac Surg. 2022; 34(4): 597-603.
Doi: 10.1093/icvts/ivab264
[OPEN ACCESS]
Web of Science
PubMed
FullText
FullText_MUG
- Co-authors Med Uni Graz
-
Zimpfer Daniel
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
- OBJECTIVES: The aim of this retrospective study was to determine if Cormatrix® (CM) represents a safe alternative to conventional patch materials used in congenital heart surgery. METHODS: A total of 57 paediatric patients who underwent cardiac surgery using an Extracellular Matrix Bioscaffold (CM) were categorized into 4 groups according to the patch implant location. Patch-related complications and reintervention rates were analysed. A subgroup of 18 patients was subsequently compared to a matched group of 36 patients who underwent similar surgical procedures with autologous pericardium as patch material. RESULTS: No patient died during hospitalization. There were 2 late deaths, not related to the implanted CM patch. Fourteen (66.7%) out of 21 patients with arterial patch plasty developed progressive vessel/right ventricular outflow tract stenosis or aneurysm. All 3 patients with a valved CM conduit developed haemodynamically relevant valve stenosis or regurgitation. A total of 18 (31.5%) patients needed reintervention and 12 (21.1%) related to CM. Four (7%) patients needed surgical treatment with operative removal of the stenosis. Redo valve replacement was performed on 2 (3.5%) patients. Six (10.5%) patients required an interventional cardiology procedure at a median interval of 5 months from surgery. The subgroup analysis revealed a significantly lower patch-related reintervention rate in patients treated with autologous pericardium when compared to CM (P = 0.006). CONCLUSIONS: CM is safe for atrial and ventricular defect closure. The use of CM for arterial vessel reconstruction is associated with higher reintervention rates when compared to autologous pericardium.
- Find related publications in this database (using NLM MeSH Indexing)
-
Cardiac Surgical Procedures - adverse effects, methods
-
Child - administration & dosage
-
Heart Defects, Congenital - diagnostic imaging, surgery
-
Heart Ventricles - administration & dosage
-
Humans - administration & dosage
-
Pericardium - administration & dosage
-
Retrospective Studies - administration & dosage
-
Treatment Outcome - administration & dosage
- Find related publications in this database (Keywords)
-
Cormatrix(R)
-
Small intestinal submucosa-extracellular matrix
-
Congenital heart defect Bioscaffold