Diagnosis and Treatment Strategies of Outflow Graft Obstruction in the Fully Magnetically Levitated Continuous-Flow centrifugal Left Ventricular Assist Device: A Multicenter Case Series.
ASAIO J. 2021; 67(1):e52-e54 Doi: 10.1097/MAT.0000000000001213 (- Case Report)
Web of Science PubMed FullText FullText_MUG

 

Co-authors Med Uni Graz
Zimpfer Daniel
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
Outflow graft obstruction in left ventricular assist devices (LVADs) is a rare complication whose clinical presentation may be subtle. We present six cases of outflow graft obstruction in patients on HeartMate 3 LVAD detected between 648 and 1,222 days on support. Detection principles are described and treatment strategies discussed. Three patients were successfully managed with stents, one underwent surgical revision, one patient died despite emergency treatment, and the last patient without symptoms was listed for heart transplantation.
Find related publications in this database (using NLM MeSH Indexing)
Adult - administration & dosage
Aged - administration & dosage
Female - administration & dosage
Heart Failure - surgery
Heart-Assist Devices - adverse effects
Humans - administration & dosage
Middle Aged - administration & dosage
Prosthesis Failure - administration & dosage
Reoperation - administration & dosage
Thrombosis - diagnosis, etiology, therapy

Find related publications in this database (Keywords)
left ventricular assist device
HeartMate 3
outflow graft obstruction
© Med Uni GrazImprint