Ghodrati, M; Maurer, A; Schlöglhofer, T; Khienwad, T; Zimpfer, D; Beitzke, D; Zonta, F; Moscato, F; Schima, H; Aigner, P.
The influence of left ventricular assist device inflow cannula position on thrombosis risk.
Artif Organs. 2020; 44(9):939-946 Doi: 10.1111/aor.13705 (- Case Report) [OPEN ACCESS]
Web of Science PubMed PUBMED Central FullText FullText_MUG

 

Co-authors Med Uni Graz
Beitzke Dietrich
Zimpfer Daniel
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
The use of left ventricular assist devices (LVADs) as a treatment method for heart failure patients has been steadily increasing; however, pathological studies showed presence of thrombi around the HeartWare ventricular assist device inflow cannula (IC) in more than 95% of patients after device explantation. Flow fields around the IC might trigger thrombus formation and require further investigation. In this study flow dynamics parameters were evaluated for different patient geometries using computational fluid dynamics (CFD) simulations. Left ventricular (LV) models of two LVAD patients were obtained from CT scans. The LV volumes of Patient 1 (P1) and Patient 2 (P2) were 264 and 114 cm3 with an IC angle of 20° and 9° from the mitral-IC tip axis at the coronal plane. The IC insertion site at the apex was central for P1, whereas it was lateral for P2. Transient CFD simulations were performed over 9 cardiac cycles. The wedge area was defined from the cannula tip to the wall of the LV apex. Mean velocity magnitude and blood stagnation region (volume with mean velocity <5 mm/s) as well as the wall shear stress (WSS) at the IC surface were calculated. Cardiac support resulted in a flow mainly crossing the ventricle from the mitral valve to the LVAD cannula for P2, while the main inflow jet deviated toward the septal wall in P1. Lower WSS at the IC surface and consequently larger stagnation volumes were observed for P2 (P1: 0.17, P2: 0.77 cm3 ). Flow fields around an LVAD cannula can be influenced by many parameters such as LV size, IC angle, and implantation site. Careful consideration of influencing parameters is essential to get reliable evaluations of the apical flow field and its connection to apical thrombus formation. Higher blood washout and lower stagnation were observed for a central implantation of the IC at the apex.
Find related publications in this database (using NLM MeSH Indexing)
Aged - administration & dosage
Cannula - adverse effects
Computer Simulation - administration & dosage
Heart Failure - physiopathology, surgery
Heart Ventricles - diagnostic imaging, physiopathology
Heart-Assist Devices - adverse effects
Hemodynamics - physiology
Humans - administration & dosage
Imaging, Three-Dimensional - administration & dosage
Male - administration & dosage
Models, Cardiovascular - administration & dosage
Stress, Mechanical - administration & dosage
Thrombosis - etiology, prevention & control
Tomography, X-Ray Computed - administration & dosage

Find related publications in this database (Keywords)
apical washout
intraventricular flow pattern
left ventricle assist device
thrombosis risk