Moscato, F; Wirrmann, C; Granegger, M; Eskandary, F; Zimpfer, D; Schima, H.
Use of continuous flow ventricular assist devices in patients with heart failure and a normal ejection fraction: a computer-simulation study.
J Thorac Cardiovasc Surg. 2013; 145(5):1352-8 Doi: 10.1016/j.jtcvs.2012.06.057
Web of Science PubMed FullText FullText_MUG

 

Co-authors Med Uni Graz
Zimpfer Daniel
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
OBJECTIVES: Continuous flow left ventricular assist devices are used in end-stage systolic heart failure. However, about one half of the patients with heart failure exhibit diastolic dysfunction with a normal ejection fraction. In the present study, the possible hemodynamic consequences of continuous flow left ventricular assist devices use for these patients were investigated. METHODS: A previously developed cardiovascular model was modified to reproduce the peculiar hemodynamics of heart failure with a normal ejection fraction. The model was based on and validated with patient data derived from the published data. A continuous flow left ventricular assist device model was included and the hemodynamic effects of pump support evaluated at rest and during exercise. RESULTS: The model accurately reproduced the published data both at rest and during exercise, leading to simulated hemodynamic values within the standard deviations of patient variability. At rest, pump support decreased the end-diastolic left ventricular pressure (6 vs 15 mm Hg) and volume (88 vs 135 mL). During exercise, maximal pump support substantially unloaded the left ventricle (end-diastolic pressure, 14 vs 35 mm Hg; volume, 133 vs 158 mL) and the pulmonary venous circulation (left atrial pressure, 12 vs 24 mm Hg) and resulted in a slight increase in cardiac output (11.7 vs 9.9 L/min). CONCLUSIONS: The simulation results suggested that continuous flow left ventricular assist devices improve the hemodynamics in patients with heart failure and a normal ejection fraction. For an optimal use of continuous flow left ventricular assist devices, low speeds should be maintained at rest, to avoid suction. However, during physical activity, higher speeds are needed to prevent an abnormal increase in the ventricular filling pressures typical of patients with heart failure and a normal ejection fraction.
Find related publications in this database (using NLM MeSH Indexing)
Atrial Function, Left - administration & dosage
Computer Simulation - administration & dosage
Diastole - administration & dosage
Exercise - administration & dosage
Heart Failure - physiopathology, therapy
Heart-Assist Devices - administration & dosage
Humans - administration & dosage
Models, Cardiovascular - administration & dosage
Prosthesis Design - administration & dosage
Reproducibility of Results - administration & dosage
Rest - administration & dosage
Stroke Volume - administration & dosage
Ventricular Function, Left - administration & dosage
Ventricular Pressure - administration & dosage