Medizinische Universität Graz - Research portal

Logo MUG Resarch Portal

Selected Publication:

SHR Neuro Cancer Cardio Lipid Metab Microb

Haberl, T; Riebandt, J; Mahr, S; Laufer, G; Rajek, A; Schima, H; Zimpfer, D.
Viennese approach to minimize the invasiveness of ventricular assist device implantation†.
Eur J Cardiothorac Surg. 2014; 46(6):991-6; discussion 996 Doi: 10.1093/ejcts/ezu051
Web of Science PubMed FullText FullText_MUG

 

Leading authors Med Uni Graz
Zimpfer Daniel
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
OBJECTIVE: Avoiding full sternotomy and cardiopulmonary bypass (CPB) could significantly reduce the invasiveness of left ventricular assist device (LVAD) implantation. Therefore, we developed minimally invasive implant strategies for the Heartware® VAD (HVAD) and the Thoratec® HeartMate II (HMII) covering isolated LVAD implantation as well as concomitant valve procedures (aortic/tricuspid). We present the surgical techniques and the initial clinical experience. METHODS: From February 2012 to March 2013, 27 patients (mean age 58 ± 8 years; male 85%; Ischemic Cardiomyopathy 63%; redo surgery 22%; Interagency Registry for Mechanically Assisted Circulatory Support Level I: 29%, II: 22%, III: 33%, IV-VII: 16%) underwent minimally invasive LVAD implantation at our department. Apical cannulation was performed via a left lateral minithoracotomy in HVAD patients (n = 20) or a left subcostal incision in HMII patients (n = 7). The outflow graft anastomosis was performed to the ascending aorta via a right minithoracotomy in the second intercostal space (n = 22) or the right subclavian artery (n = 2). If additional valve procedures (aortic/tricuspid) were necessary (n = 3), a hemisternotomy was performed to access the valve and perform the outflow graft anastomosis. Circulatory support for LVAD implantation was CPB (33%), extracorporeal membrane oxygenation (48%) or off-pump (19%). RESULTS: The minimally invasive approach was feasible in all patients with no need for conversions. Thirty-day and in-hospital mortality were 7.4 and 14.8%, respectively. In-hospital stay was 30.0 ± 22.5 days. One patient (4%) died during follow-up from pump thrombus formation. Three patients (11%) underwent surgical revision for bleeding. CONCLUSIONS: Minimally invasive LVAD implantation is feasible and safe. The very encouraging results obtained in this initial series justify a broad application of this technique.
Find related publications in this database (using NLM MeSH Indexing)
Aged - administration & dosage
Austria - administration & dosage
Cardiomyopathies - surgery
Female - administration & dosage
Heart-Assist Devices - administration & dosage
Humans - administration & dosage
Male - administration & dosage
Middle Aged - administration & dosage
Minimally Invasive Surgical Procedures - adverse effects, methods
Myocardial Ischemia - surgery
Prosthesis Implantation - adverse effects, methods
Retrospective Studies - administration & dosage

Find related publications in this database (Keywords)
Ventricular assist device
Minimal invasive implantation
© Med Uni GrazImprint