Medizinische Universität Graz - Research portal

Logo MUG Resarch Portal

Selected Publication:

SHR Neuro Cancer Cardio Lipid Metab Microb

Mert, A; Gan, LS; Knosp, E; Sutherland, GR; Wolfsberger, S.
Advanced cranial navigation.
Neurosurgery. 2013; 72 Suppl 1:43-53 Doi: 10.1227/NEU.0b013e3182750c03
Web of Science PubMed FullText FullText_MUG

 

Leading authors Med Uni Graz
Wolfsberger Stefan
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
BACKGROUND: Cranial surgical navigation is most commonly performed by registration with fiducial markers, optic tracking, and intermittent pointer-based application. OBJECTIVE: To assess the accuracy and applicability of an advanced cranial navigation setup. METHODS: Continuous electromagnetic instrument navigation was used in 136 neurosurgical cases with a standard navigation system. A phantom head in an intraoperative magnetic resonance imaging environment was used to compare the accuracy of the advanced and standard navigation setups. RESULTS: A navigated suction device was used in 71 cases of intracranial tumor surgery and 46 cases of endoscopic transsphenoidal surgery. The ventriculoscope was navigated in 6 cases and the stereotactic biopsy needle in 4 cases. Electromagnetic tracking was used for catheter placement in 9 cases. The learning curve comprised 6 of the 136 cases during the first month of application. No significant difference was observed at the intracranial target points between the standard navigation setup using optic tracking, fiducial marker registration, and pointer and the advanced navigation setup with electromagnetic tracking, surface-based registration, and navigation of a field-detecting stylet in a standard metal suction tube when performed outside the 5-G line of the 3.0-T intraoperative magnetic resonance imaging. CONCLUSION: Continuous instrument navigation is the prerequisite for seamless integration of navigation systems into the neurosurgical operating workflow. Our data confirm that the application of preoperative imaging, surface-merge registration, and continuous electromagnetic tip-tracked instrument navigation may provide such integration without a significant reduction in accuracy compared with standard navigation.
Find related publications in this database (using NLM MeSH Indexing)
Arachnoid Cysts - pathology, surgery
Biopsy, Needle - administration & dosage
Brain Neoplasms - pathology, surgery
Consciousness - administration & dosage
Ergonomics - administration & dosage
Fiducial Markers - administration & dosage
Head Movements - administration & dosage
Humans - administration & dosage
Magnetic Resonance Imaging - methods, standards
Magnets - administration & dosage
Monitoring, Intraoperative - methods, standards
Neuroendoscopy - instrumentation, methods
Neuronavigation - methods, standards
Phantoms, Imaging - administration & dosage
Polyurethanes - administration & dosage
Preoperative Care - methods, standards
Reproducibility of Results - administration & dosage
Retrospective Studies - administration & dosage
Sphenoid Bone - surgery
Task Performance and Analysis - administration & dosage
Tomography, X-Ray Computed - methods, standards

Find related publications in this database (Keywords)
Cranial navigation
Electromagnetic tracking
Instrument tracking
Surface registration