Kron, S; Schneditz, D; Leimbach, T; Kron, J.
Feedback control of absolute blood volume: A new technical approach in hemodialysis.
Hemodial Int. 2020; 10(1): Doi: 10.1111/hdi.12826
Web of Science PubMed FullText FullText_MUG

 

Co-authors Med Uni Graz
Schneditz Daniel
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
The success of automatic feedback control systems to improve hemodynamic stability by preventing relative blood volume from dropping beyond a critical value during dialysis is limited. The aim of this study was to use one of these systems for control of absolute rather than relative blood volume to prevent volume-dependent morbid events. Dialysis was delivered by a machine providing feedback control of ultrafiltration rates, relative blood volume monitoring, and accurate bolus infusion of 240 mL of dialysate to measure absolute blood volume at the beginning of dialysis. Critical relative blood volume required by the control algorithm was calculated from absolute blood volume at the beginning and a critical absolute blood volume of 67 mL/kg. In 40 stable patients, ultrafiltration was guided by blood volume using the feedback algorithm of the integrated program. Blood volume was maintained in a narrow range above the prespecified minimal value of 67 mL/kg. At the end of dialysis, absolute blood volume ranged from 67.5 to 72.5 mL/kg (69.4 ± 1.3 mL/kg). No volume-dependent intradialytic morbid event occurred. A feedback control system for relative blood volume-controlled ultrafiltration can be used for control of absolute blood volume. A prescribed target of absolute blood volume can be converted into relative blood volume, and this can subsequently be reached automatically with the integrated feedback system of the dialysis machine. Intradialytic morbid events could be considerably reduced. The whole procedure could be completely automated without altering the hardware of the dialysis device. © 2020 International Society for Hemodialysis.

Find related publications in this database (Keywords)
Absolute blood volume
target blood volume
automated volume management
blood volume monitoring
chronic hemodialysis
© Med Uni GrazImprint