Pea, F; Krause, R; Müller, C; Hennart, B; Richardson, M; Meinitzer, A; Wiesen, MHJ; Wiktorowicz, T; Spickermann, J; Henriksen, AS.
Interlaboratory Analysis of Isavuconazole Plasma Concentration Assays Among European Laboratories.
Ther Drug Monit. 2019; 41(5):657-664 Doi: 10.1097/FTD.0000000000000632 [OPEN ACCESS]
Web of Science PubMed PUBMED Central FullText FullText_MUG

 

Co-Autor*innen der Med Uni Graz
Krause Robert
Meinitzer Andreas
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
Under certain circumstances, clinicians treating patients with isavuconazole for invasive aspergillosis or mucormycosis may use therapeutic drug monitoring. However, the accuracy and reproducibility of the various assays used by different laboratories for the quantification of isavuconazole plasma concentrations have yet to be determined. Human plasma samples spiked with known concentrations of isavuconazole were provided to 27 European laboratories that took part in a "round-robin" test (an interlaboratory test performed independently at least 2 times; 2 rounds performed in the current study). Assay methods included liquid chromatography-tandem mass spectrometry (LC-MS/MS), LC with ultraviolet detection (LC-UV), LC with fluorescence detection (LC-FL), and bioassay. The accuracy and reproducibility compared with the known concentrations for each sample in each round were compared overall, between assays, and between laboratories. Twenty-seven laboratories participated in the study (LC-MS/MS, n = 15; LC-UV; n = 9; LC-FL, n = 1; bioassay, n = 2). In round 1, for nominal concentrations of 1000, 1700, 2500, and 4000 ng/mL, the mean (SD) determined concentrations were 1007 (183), 1710 (323), 2528 (540), and 3898 (842) ng/mL, respectively. In round 2, for nominal concentrations of 1200, 1800, 2400, and 4000 ng/mL, the mean (SD) determined concentrations were 1411 (303), 2111 (409), 2789 (511), and 4723 (798) ng/mL, respectively. Over both rounds, determined concentrations were consistently within 15% of the nominal concentrations for 10 laboratories (LC-MS/MS, n = 4; LC-UV, n = 5; bioassay, n = 1) and consistently exceeded the upper 15% margin for 7 laboratories (LC-MS/MS and LC-UV, n = 3 each; LC-FL, n = 1). Alignment of methodologies among laboratories may be warranted to improve the accuracy and reproducibility of therapeutic drug measurements.
Find related publications in this database (using NLM MeSH Indexing)
Biological Assay - methods
Chromatography, Liquid - methods
Drug Monitoring - methods
Drug Monitoring -
Humans -
Laboratories -
Nitriles - blood
Plasma - chemistry
Pyridines - blood
Reproducibility of Results -
Tandem Mass Spectrometry - methods
Triazoles - blood