Enko, D; Stelzer, I; Böckl, M; Derler, B; Schnedl, WJ; Anderssohn, P; Meinitzer, A; Herrmann, M.
Comparison of the diagnostic performance of two automated urine sediment analyzers with manual phase-contrast microscopy.
Clin Chem Lab Med. 2020; 58(2):268-273 Doi: 10.1515/cclm-2019-0919
Web of Science PubMed FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Enko Dietmar
Co-Autor*innen der Med Uni Graz
Herrmann Markus
Meinitzer Andreas
Schnedl Wolfgang
Stelzer Ingeborg
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
Background Recently, several manufacturers have launched automated urinalysis platforms. This study aimed to compare the diagnostic performance of the UF-5000 (Sysmex Corporation, Kobe, Japan) and the cobas® u 701 (Roche Diagnostics, Rotkreuz, Switzerland) urine sediment analyzers with manual phase-contrast microscopy as the reference method. Methods A total of 195 urine samples were analyzed on both automated platforms and subjected to manual microscopic examination. Agreement was assessed by Cohen's kappa (κ) analysis. Sensitivities and specificities were calculated. Results The agreement of the UF-5000 with manual microscopy was almost perfect (κ > 0.8) for red (RBC) and white blood cells (WBC), renal tubular epithel cells, hyaline casts, bacteria (BACT) and yeast (YLC), substantial (κ = 0.61-0.80) for squamous epithel cells (SEC) and pathologic casts, and moderate (κ = 0.41-0.60) for transitional epithel cells. The cobas® u 701 showed substantial agreement (κ = 0.61-0.80) for WBC, moderate agreement (κ = 0.41-0.60) for hyaline casts, and fair agreement (κ = 0.21-0.40) for RBC, SEC, non-squamous epithel (NEC), pathologic casts, BACT and YLC. The UF-5000 sensitivities ranged between 98.5% for RBC and 83.3% for pathological casts. The cobas® u 701 showed sensitivities between 83.0% for WBC and 31.6% for YLC. Conclusions The UF-5000 (Sysmex) analyzer showed a better diagnostic agreement with manual phase-contrast microscopy compared to the cobas® u 701 (Roche) module. The Sysmex platform showed reliable results for urine sediment analysis. However, pathological samples should be verified with manual microscopy.

Find related publications in this database (Keywords)
automation
microscopy
urinalysis
urine sediment
© Med Uni Graz Impressum