Grunert, R; Schleifenbaum, S; Möbius, R; Kopper, M; Rotsch, C; Drossel, WG; Hammer, N; Prietzel, T.
Novel concept of a modular hip implant could contribute to less implant failure in THA: a hypothesis.
Patient Saf Surg. 2018; 12(1):1-1
Doi: 10.1186/s13037-017-0148-7
[OPEN ACCESS]
Web of Science
PubMed
FullText
FullText_MUG
- Co-Autor*innen der Med Uni Graz
-
Hammer Niels
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
-
The modularity in total hip arthroplasty (THA) allows orthopaedic surgeons for an exact reconstruction of hip biomechanical parameters especially in revision and tumor arthroplasty. Modular structured femoral stems using taper junctions showed increased implant breakage in the recent past.
We hypothesize that a novel modular stem-neck-interface leads to less implant breakage compared to conventional femoral stems.
For this purpose, a novel modular femoral stem for THA was to design and manufacture. Therefore, three different variants of interface mechanisms were developed that enable a simple connection between the stem and the neck modules and allow for intra-operatively adjustment. Three prototypes A, B and C were manufactured and subsequently dynamic fatigue (ISO 7206-6) and body donor tested.
Modularity in THA is mainly applied in THA as well as in revision and tumor arthroplasty. Modular implants are barely used because of the high risk of breakage. Another risks in this context are taper fretting, corrosion and disconnection. With the novel design, it should be possible to detach the stem and neck module intra-operatively to adapt the anatomical situation. The novel coupling mechanism of the rotating interface seems to be the most suitable for a secure stem-neck connection and is characterized by good intraoperative handling.