Dörr, M; Richter, S; Eckert, S; Ohlow, MA; Hammer, F; Hummel, A; Dornberger, V; Genzel, E; Baulmann, J.
Invasive Validation of Antares, a New Algorithm to Calculate Central Blood Pressure from Oscillometric Upper Arm Pulse Waves.
J CLIN MED. 2019; 8(7):
Doi: 10.3390/jcm8071073
[OPEN ACCESS]
Web of Science
PubMed
FullText
FullText_MUG
- Leading authors Med Uni Graz
-
Baulmann Johannes
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
-
Background:
Antares is an algorithm for pulse wave analysis (PWA) by oscillometric blood pressure (BP) monitors in order to estimate central (aortic) blood pressure (cBP). Antares aims to enable brachial cuff-based BP monitors to be type II-devices, determining absolute cBP values independently of potential peripheral BP inaccuracies. The present study is an invasive validation of the Antares algorithm in the custo screen 400.
Methods:
We followed entirely the 2017 ARTERY protocol for validation of non-invasive cBP devices, the 2013 American National Standards Institute, Inc./Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2 and 2018 AAMI/European Society of Hypertension (ESH)/ISO validation standard protocols. In total, 191 patients undergoing cardiac catheterization were included, of which 145 patients entered analysis. Invasive cBP recordings were compared to simultaneous non-invasive cBP estimations using the Antares algorithm, integrated into an oscillometric BP monitor.
Results:
Mean difference between invasive and non-invasively estimated systolic cBP was 0.71 mmHg with standard deviation of 5.95 mmHg, fulfilling highest validation criteria.
Conclusion:
Antares is the first algorithm for estimation of cBP that entirely fulfills the 2017 ARTERY and AAMI/ESH/ISO validation protocols. The Antares algorithm turns the custo screen 400 BP monitor into a type II-device. Integration of Antares into commercially available BP monitors could make it possible to measure PWA parameters in virtually every practice in future.
- Find related publications in this database (Keywords)
-
validation
-
invasive
-
central blood pressure
-
pulse wave analysis
-
antares