SHR Neuro Krebs Kardio Lipid Stoffw Microb

Freude, T; Schroeter, S; Plecko, M; Bahrs, C; Martetschlaeger, F; Kraus, TM; Stoeckle, U; Doebele, S.
Dynamic-locking-screw (DLS)-leads to less secondary screw perforations in proximal humerus fractures.
BMC Musculoskelet Disord. 2014; 15(5): 194-194. Doi: 10.1186/1471-2474-15-194 [OPEN ACCESS]
Web of Science PubMed PUBMED Central FullText FullText_MUG

 

Co-Autor*innen der Med Uni Graz
Plecko Michael
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
Loss of reduction and screw perforation causes high failure rates in the treatment of proximal humerus fractures. The purpose of the present study was to evaluate the early postoperative complications using modern Dynamic Locking Screws (DLS 3.7) for plating of proximal humerus fractures. Between 03/2009 and 12/2010, 64 patients with acute proximal humerus fractures were treated by angular stable plate fixation using DLSs in a limited multi-centre study. Follow-up examinations were performed three, six, twelve and twenty-four weeks postoperatively and any complications were carefully collected. 56 of 64 patients were examined at the six-month follow-up. Complications were observed in 12 patients (22%). In five cases (9%), a perforation of the DLS 3.7 occurred. Despite the use of modern DLS 3.7, the early complications after plating of proximal humerus fractures remain high. The potential advantage of the DLS 3.7 regarding secondary screw perforation has to be confirmed by future randomized controlled trials.
Find related publications in this database (using NLM MeSH Indexing)
Aged -
Bone Plates -
Bone Screws - adverse effects
Equipment Design -
Female -
Follow-Up Studies -
Fracture Fixation, Internal - instrumentation
Fracture Healing -
Humans -
Male -
Middle Aged -
Osteonecrosis - etiology
Postoperative Complications - etiology
Shoulder Fractures - surgery
Trauma Centers -