A head-to-head comparison of personal and professional continuous glucose monitoring systems in people with type 1 diabetes: Hypoglycaemia remains the weak spot.
Diabetes Obes Metab. 2019; 21(4):1043-1048 Doi: 10.1111/dom.13598 [OPEN ACCESS]
Web of Science PubMed PUBMED Central FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Moser Othmar
Pandis Marlene
Sourij Harald
Co-Autor*innen der Med Uni Graz
Aberer Felix
Elsayed Hesham
Hochfellner Daniel
Kojzar Harald
Mader Julia
Pieber Thomas
Wimmer Melanie
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
To compare the performance of a professional continuous glucose monitoring (proCGM) and a personal continuous glucose monitoring (persCGM) system worn in parallel under standardized conditions in individuals with type 1 diabetes (T1D), two CGM systems (iPro2 - proCGM; Minimed 640G - persCGM) worn in parallel using the same sensor (Enlite 2) were compared. Ten people with T1D were included in this single-centre, open-label study in which CGM performance was evaluated. The study consisted of a 24-hours inpatient phase (meals, exercise, glycaemic challenges) and a 4-day home phase. Analyses included fulfilment of ISO 15197:2013 criteria, mean absolute relative difference (MARD), Parkes Error Grid and Bland-Altman plots. During the inpatient stay, ISO 15197:2013 criteria fulfilment was 58.4% (proCGM) and 57.8% (persCGM). At home, the systems met ISO 15197:2013 criteria by 66.5% (proCGM) and 65.3% (persCGM). No difference of MARD in inpatient phase (19.1 ± 16.7% vs. 19.0 ± 19.6; P = 0.83) and home phase (18.6 ± 26.8% vs. 17.4 ± 21.3%, P = 0.87) was observed. All sensors performed less accurately during hypoglycaemia. ProCGM and persCGM showed similar performance during daytime and night-time for the inpatient and the home phase. However, sensor performance was reduced during hypoglycaemia for both systems. © 2018 The Authors. Diabetes, Obesity and Metabolism published by John Wiley & Sons Ltd.
Find related publications in this database (using NLM MeSH Indexing)
Adult -
Blood Glucose - metabolism
Blood Glucose Self-Monitoring - instrumentation
Diabetes Mellitus, Type 1 - drug therapy
Diabetes Mellitus, Type 1 - metabolism
Female -
Humans -
Hyperglycemia - diagnosis
Hyperglycemia - metabolism
Hypoglycemia - diagnosis
Hypoglycemia - metabolism
Hypoglycemic Agents - therapeutic use
Insulin - therapeutic use
Male -
Monitoring, Ambulatory - instrumentation
Monitoring, Physiologic - instrumentation
Young Adult -

Find related publications in this database (Keywords)
clinical trial
continuous glucose monitoring (CGM)
hypoglycaemia
type 1 diabetes
© Med Uni Graz Impressum