Medizinische Universität Graz - Research portal

Logo MUG Resarch Portal

Selected Publication:

Dahaba, AA; Von Klobucar, F; Rehak, PH; List, WF.
Comparison of a new piezoelectric train-of-four neuromuscular monitor, the ParaGraph, and the Relaxometer mechanomyograph.
Br J Anaesth. 1999; 82(5):780-782 Doi: 10.1093/bja/82.5.780 [OPEN ACCESS]
Web of Science PubMed FullText FullText_MUG Google Scholar

 

Co-authors Med Uni Graz
Rehak Peter
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
The ParaGraph is a new device for monitoring neuromuscular function using a piezoelectric motion sensor. In 20 patients, monitoring of neuromuscular block produced by cisatracurium 0.1 mg kg-1 was compared using the ParaGraph and a Relaxometer 2 mechanomyograph. The ParaGraph was quick to set up, and easy to operate and interpret. There were no significant differences in the time to 100% depression of T1/T0, time to 25% recovery of T1/T0 or time to recovery of T1/T0 from 25% to 75%, measured by the two monitors. When the difference between the two monitors was plotted against the average of the two measurements, the limits of agreement for T1/T0 (-42.95, +53.98%) and the train-of-four ratio, T4/T1 (-0.28, +0.21) were too wide to allow the values given by the two monitors for individual patients to be used interchangeably.
Find related publications in this database (using NLM MeSH Indexing)
Adolescent -
Adult -
Atracurium - analogs and derivatives
Comparative Study - analogs and derivatives
Electromyography - analogs and derivatives
Female - analogs and derivatives
Humans - analogs and derivatives
Male - analogs and derivatives
Middle Aged - analogs and derivatives
Monitoring, Intraoperative - instrumentation
Neuromuscular Blockade - instrumentation
Neuromuscular Junction - drug effects
Neuromuscular Nondepolarizing Agents - pharmacology

© Med Uni GrazImprint