Medizinische Universität Graz Austria/Österreich - Forschungsportal - Medical University of Graz

Logo MUG-Forschungsportal

Gewählte Publikation:

SHR Neuro Krebs Kardio Lipid Stoffw Microb

Sadoghi, P; Hasenhütl, S; Gruber, G; Leitner, L; Leithner, A; Rumpold-Seitlinger, G; Kastner, N; Poolman, RW; Glehr, M.
Impact of a new cryotherapy device on early rehabilitation after primary total knee arthroplasty (TKA): a prospective randomised controlled trial.
Int Orthop. 2018; 42(6):1265-1273 Doi: 10.1007/s00264-018-3766-5 [OPEN ACCESS]
Web of Science PubMed FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Gruber Gerald
Sadoghi Patrick
Co-Autor*innen der Med Uni Graz
Glehr Mathias
Kastner Norbert
Leithner Andreas
Leitner Lukas
Rumpold-Seitlinger Gudrun
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
The aim of this prospective, randomised and single blinded study was to evaluate the efficiency and safety of a new cryotherapy device in patients undergoing unilateral, primary total knee arthroplasty (TKA). Our hypothesis was that patients administered to the new cryotherapy device would perform better than patients receiving a conventional standard cold therapy regimen. Ninety-seven patients were randomised into two groups receiving either the cTreatment® (new cryotherapy device) or the standard cold therapy protocol (including cold pack application for six days after the surgical intervention). We evaluated the following endpoints consisting of range of motion (ROM), pain intensity, and knee girth on admission day and the second, fourth, and sixth post-operative day (POD). A statistically significant benefit of the new cryotherapy device was detected regarding the ROM on the sixth POD with an average gain of 7 degrees (p = 0.021). Pain in the numeric rating scale (NRS) score in motion was significantly lower in the cTreatment® group on the second POD (p = 0.034). There were no statistically significant differences between groups regarding the NRS in rest, patient controlled analgesia (PCA) consumption, and girth measurements. No adverse effects were observed in both study groups. The new computer-controlled cooling therapy device provides benefits in terms of early post-operative remobilisation with respect to ROM and pain, which might be attributed to a reduced inflammatory response, as well as reduced secretion and bleeding. The cTreatment® system appears to be a safe and efficient procedure.

Find related publications in this database (Keywords)
Total knee arthroplasty
Cryotherapy
Fast-track arthroplasty
Pain relief
Prospective randomised trial
© Med Uni Graz Impressum