Manninger, M; Odeneg, T; Fruhwald, F; Brussee, H; Scherr, D.
Oversensing of the wearable cardioverter defibrillator during bipolar ventricular stimulation.
Wien Klin Wochenschr. 2017; 129(23-24):910-912 Doi: 10.1007/s00508-017-1290-z [OPEN ACCESS]
Web of Science PubMed PUBMED Central FullText FullText_MUG

 

Leading authors Med Uni Graz
Manninger-Wünscher Martin
Co-authors Med Uni Graz
Brussee Helmut
Fruhwald Friedrich
Scherr Daniel
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
The wearable cardioverter defibrillator (WCD) is a temporary treatment option for patients with potentially reversible risk of sudden cardiac death. This case demonstrates a pitfall during WCD usage in a pacemaker-dependent patient as well as a possible solution allowing continuation of WCD therapy. Bipolar stimulation may lead to double counting of the WCD detection algorithm resulting in false alarm or inappropriate therapy.