Farrell, CJ; Soldo, J; McWhinney, B; Bandodkar, S; Herrmann, M.
Impact of assay design on test performance: lessons learned from 25-hydroxyvitamin D.
Clin Chem Lab Med. 2014; 52(11):1579-1587 Doi: 10.1515/cclm-2014-0111
Web of Science PubMed FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Herrmann Markus
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
Current automated immunoassays vary significantly in many aspects of their design. This study sought to establish if the theoretical advantages and disadvantages associated with different design formats of automated 25-hydroxyvitamin D (25-OHD) assays are translated into variations in assay performance in practice. 25-OHD was measured in 1236 samples using automated assays from Abbott, DiaSorin, Roche and Siemens. A subset of 362 samples had up to three liquid chromatography-tandem mass spectrometry 25-OHD analyses performed. 25-OHD₂ recovery, dilution recovery, human anti-animal antibody (HAAA) interference, 3-epi-25-OHD₃ cross-reactivity and precision of the automated assays were evaluated. The assay that combined release of 25-OHD with analyte capture in a single step showed the most accurate 25-OHD₂ recovery and the best dilution recovery. The use of vitamin D binding protein (DBP) as the capture moiety was associated with 25-OHD₂ under-recovery, a trend consistent with 3-epi-25-OHD₃ cross-reactivity and immunity to HAAA interference. Assays using animal-derived antibodies did not show 3-epi-25-OHD₃ cross-reactivity but were variably susceptible to HAAA interference. Not combining 25-OHD release and capture in one step and use of biotin-streptavidin interaction for solid phase separation were features of the assays with inferior accuracy for diluted samples. The assays that used a backfill assay format showed the best precision at high concentrations but this design did not guarantee precision at low 25-OHD concentrations. Variations in design among automated 25-OHD assays influence their performance characteristics. Consideration of the details of assay design is therefore important when selecting and validating new assays.
Find related publications in this database (using NLM MeSH Indexing)
25-Hydroxyvitamin D 2 - blood
25-Hydroxyvitamin D 2 - immunology
25-Hydroxyvitamin D 2 - isolation & purification
Antibodies - immunology
Automation -
Chromatography, High Pressure Liquid -
Cross Reactions -
Humans -
Immunoassay - instrumentation
Immunoassay - methods
Protein Binding -
Tandem Mass Spectrometry -
Vitamin D - analogs & derivatives
Vitamin D - blood
Vitamin D - immunology
Vitamin D - isolation & purification
Vitamin D-Binding Protein - chemistry
Vitamin D-Binding Protein - metabolism

Find related publications in this database (Keywords)
25-hydroxyvitamin D
immunoassay
liquid chromatography-tandem mass spectrometry
© Med Uni Graz Impressum