Seule, M; Sikorski, C; Sakowitz, O; von Campe, G; Santos, E; Orakcioglu, B; Unterberg, A; Keller, E.
Evaluation of a New Brain Tissue Probe for Intracranial Pressure, Temperature, and Cerebral Blood Flow Monitoring in Patients with Aneurysmal Subarachnoid Hemorrhage.
Neurocrit Care. 2016; 25(2):193-200
Doi: 10.1007/s12028-016-0284-4
[OPEN ACCESS]
Web of Science
PubMed
FullText
FullText_MUG
- Co-Autor*innen der Med Uni Graz
-
von Campe Gord
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
-
To evaluate an intraparenchymal probe for intracranial pressure (ICP) and temperature (TEMP) monitoring as well as determination of cerebral hemodynamics using a near-infrared spectroscopy (NIRS) and indocyanine green (ICG) dye dilution method (NIRS-ICP probe).
The NIRS-ICP probe was applied after aneurysmal subarachnoid hemorrhage if multimodal monitoring was established due to poor neurological condition. ICP and TEMP values were obtained from ventricular catheters and systemic temperature sensors. Repeated NIRS-ICG measurements (2 injections within 30 min) were performed daily for determination of cerebral blood flow (CBF), cerebral blood volume (CBV), and mean transit time of ICG (mttICG). Secondary neurologic dysfunction was defined as brain tissue oxygen tension <20 mmHg and/or lactate/pyruvate ratio >35 obtained from cerebral probing.
A total of 128 NIRS-ICG measurements were performed in ten patients. The correlation coefficients between ICP and TEMP values obtained with the NIRS-ICP probe and values from routine monitoring were r = 0.72 and r = 0.96, respectively. The mean values were 30.3 ± 13.6 ml/100 g/min for CBF, 3.3 ± 1.2 ml/100 g for CBV, and 6.8 ± 1.6 s for mttICG. The coefficients of variation from repeated NIRS-ICG measurements were 10.9 % for CBF, 11.7 % for CBV, and 3.8 % for mttICG. The sensitivity for detection of secondary neurologic dysfunction was 85 % and the specificity 83 % using a CBF-threshold of 25 ml/100 g/min.
Multimodal monitoring using the NIRS-ICP probe is feasible with high reproducibility of measurement values and the ability to detect secondary neurologic dysfunction. No safety concerns exist for the routine clinical use of the NIRS-ICP probe.
- Find related publications in this database (using NLM MeSH Indexing)
-
Adult -
-
Body Temperature -
-
Cerebrovascular Circulation -
-
Female -
-
Humans -
-
Intracranial Aneurysm - complications
-
Intracranial Pressure -
-
Male -
-
Middle Aged -
-
Neurophysiological Monitoring - standards
-
Spectroscopy, Near-Infrared - standards
-
Subarachnoid Hemorrhage - diagnosis
-
Subarachnoid Hemorrhage - etiology
-
Subarachnoid Hemorrhage - therapy
- Find related publications in this database (Keywords)
-
Neuromonitoring
-
Secondary neurologic dysfunction
-
Near-infrared spectroscopy
-
Indocyanine green
-
Cerebral blood flow
-