Schalamon, J; Petnehazy, T; Ainoedhofer, H; Castellani, C; Till, H; Singer, G.
Experimental comparison of abdominal drainage systems.
Am J Surg. 2017; 213(6):1038-1041 Doi: 10.1016/j.amjsurg.2016.09.043
Web of Science PubMed FullText FullText_MUG

 

Leading authors Med Uni Graz
Schalamon Johannes
Singer Georg
Co-authors Med Uni Graz
Ainödhofer Herwig
Castellani Christoph
Till Holger
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
In an experimental study the performance of different closed abdominal drainage systems was tested. A vacuum bottle designed for Redon Drainage, a flexible plastic bulb designed for Jackson-Pratt drains and a V.A.C.® Negative Pressure Wound Therapy System were used. In a porcine cadaveric study mimicking the abdominal cavity the intrinsic pressure (IP) at one and three minutes (T0, T3) and the amount of evacuated fluid were measured. The Redon and Jackson drainage displayed a rapid decline to IP values of almost zero comparing T0 and T3. Only the V.A.C.® system was able to preserve constant values of negative IP values measured at both time points. Only the V.A.C.® system was able to remove almost the whole amount of inserted fluid. In an experimental setting the V.A.C.® system was superior to the other two tested systems in delivering constant negative IP and the amount of evacuated fluid. Copyright © 2016 Elsevier Inc. All rights reserved.
Find related publications in this database (using NLM MeSH Indexing)
Abdominal Cavity - surgery
Animals -
Drainage - instrumentation
Equipment Design -
Negative-Pressure Wound Therapy - instrumentation
Postoperative Complications -
Swine -

Find related publications in this database (Keywords)
Abdominal drainage
Experimental
Redon
Jackson
VAC
© Med Uni GrazImprint