Boettcher, C; Dost, A; Wudy, SA; Flechtner-Mors, M; Borkenstein, M; Schiel, R; Weitzel, D; Bechtold-Dalla Pozza, S; Wolf, J; Holl, RW; German/Austrian Diabetes Prospective Documentation Initiative.
Accuracy of blood glucose meters for self-monitoring affects glucose control and hypoglycemia rate in children and adolescents with type 1 diabetes.
Diabetes Technol Ther. 2015; 17(4):275-282
Doi: 10.1089/dia.2014.0262
Web of Science
PubMed
FullText
FullText_MUG
- Co-authors Med Uni Graz
-
Borkenstein Helmuth Martin
- Altmetrics:
- Dimensions Citations:
- Plum Analytics:
- Scite (citation analytics):
- Abstract:
-
This study investigated the accuracy of blood glucose meters for self-monitoring and its influence on glycated hemoglobin (HbA1c) levels and the frequency of hypoglycemic coma.
Self-measured and simultaneously obtained laboratory blood glucose values from 9,163 patients with type 1 diabetes <18 years of age in the German/Austrian Diabetes Prospective Documentation Initiative registry were analyzed by investigating their compliance with the International Organization for Standardization (ISO) criteria (versions 2003 and 2013) and by error grid analyses. Regression models elucidated effects on glucose control and hypoglycemia rates.
Depending on the respective subgroup (defined by sex, age, duration of diabetes, mode of insulin therapy), 78.7-94.7% of the self-monitoring of blood glucose (SMBG) values met the old and 79.7-88.6% met the new ISO criteria. In Clarke and Parkes error grid analyses, the percentages of SMBG values in Zone A ranged between 92.8% and 94.6% (Clarke) and between 92.2% and 95.0% (Parkes). The patient group with SMBG devices measuring "far too low" (compared with the laboratory-obtained glucose levels) presented with a higher HbA1c level than those measuring "far too high," "too high," "identical/almost identical," or "too low" (based on quintiles of deviation). Performing "far too high" was associated with the highest rate of hypoglycemic coma in comparison with the other deviation quintiles.
This study showed that current SMBG devices fulfilled neither the previous nor the new ISO criteria. Large deviations of the SMBG values from the "true" glucose levels resulted in higher HbA1c levels and markedly increased rates of hypoglycemic events.
- Find related publications in this database (using NLM MeSH Indexing)
-
Adolescent -
-
Austria -
-
Blood Glucose Self-Monitoring - instrumentation
-
Blood Glucose Self-Monitoring - standards
-
Child -
-
Child, Preschool -
-
Data Accuracy -
-
Databases, Factual -
-
Diabetes Mellitus, Type 1 - blood
-
Diabetes Mellitus, Type 1 - complications
-
Diabetic Coma - epidemiology
-
Female -
-
Germany -
-
Glycated Hemoglobin A - analysis
-
Humans -
-
Hypoglycemia - epidemiology
-
Infant -
-
Male -
-
Prospective Studies -
-
Reference Values -