Acta Diabetol. 2014; 51(5):883-886 Doi: 10.1007/s00592-014-0578-y
Web of Science PubMed FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Sinner Frank Michael
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
The aim of this study was to develop a single-port system that combines glucose monitoring and insulin infusion. In a preclinical proof-of-concept trial, the performance of the glucose measurement at the site of insulin infusion was assessed. Glucose levels were clamped from 40 to 250 mg/dL by intravenous glucose infusion and subcutaneous insulin infusion via the glucose sensor. Sensor-glucose values correlated well with reference blood-glucose values, despite infusion at the site of glucose measurement. The average median ARE value was 21.6 ± 5.7 % for sensors used for insulin infusion, 18.1 ± 5.8 % for sensors used for NaCl infusion and 19.2 ± 7.9 % for sensors without infusion. These preclinical in vivo results demonstrate that single-port glucose monitoring is feasible at the site of insulin infusion.
Find related publications in this database (using NLM MeSH Indexing)
Blood Glucose - analysis
Blood Glucose Self-Monitoring - instrumentation
Blood Glucose Self-Monitoring - methods
Drug Delivery Systems - instrumentation
Drug Delivery Systems - methods
Humans -
Insulin - administration & dosage
Insulin Infusion Systems -

Find related publications in this database (Keywords)
Continuous glucose monitoring
Single-port system
Optical glucose sensor
Insulin infusion
Preclinical trial