M Buckland, D; Sadoghi, P; Wimmer, MD; Vavken, P; Pagenstert, GI; Valderrabano, V; Rosso, C.
Meta-analysis on biomechanical properties of meniscus repairs: are devices better than sutures?
Knee Surg Sports Traumatol Arthrosc. 2015; 23(1):83-89 Doi: 10.1007/s00167-014-2966-9
Web of Science PubMed FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Sadoghi Patrick
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
Meniscal repair devices have been extensively tested during the past decades as reported in the literature. Reviewing the different meniscal repair devices and sutures with their respective biomechanical properties. For this meta-analysis, we conducted a systematic online search using PubMed, EMBASE, CCTR, and CINAHL using the search terms Meniscus OR Meniscal AND Biomechanics AND Repair). Load-to-failure (LtF), stiffness, and cyclic outcome measures were extracted independently and in duplicate. The systematic search revealed 841 manuscripts in total. After exclusion of duplicates and irrelevant publications, 41 studies remained for final analysis. The studies were published in English and German from 1995 to 2013. Due to differing cyclic force protocols, cyclic outcomes had to be excluded. Overall, sutures had a higher LtF [suture: 87.7 ± 0.3 N (weighted mean ± standard error), device: 56.3 ± 0.1 N] and stiffness (suture: 8.9 ± 0.04 N/mm, device: 8.6 ± 0.04 N/mm) than devices, both p < 0.05. In LfT testing, PDS 0 Vertical (145.0 ± 8.1 N), OrthoCord 2-0 (143.6 ± 11.3 N), and Ethibond No 0 Vertical (133.4 ± 7.7 N) were the strongest sutures and Meniscal Viper (140.9 ± 5.1 N), MaxFire Vertical (136.2 ± 11.3 N), and FasT-Fix Vertical (115.2 ± 1.6 N) were the strongest devices. Second-generation devices were significantly stronger and stiffer than first-generation devices (p < 0.001). Suture repair remains the gold standard with a vertically oriented suture configuration showing superior LtF values compared to a horizontal configuration. Nevertheless, some meniscal repair devices have similar biomechanical properties to suture repairs. Both suture repairs and devices have a place in meniscal restoration. None, meta-analysis of controlled laboratory studies.
Find related publications in this database (using NLM MeSH Indexing)
Absorbable Implants -
Biomechanical Phenomena -
Humans -
Menisci, Tibial - surgery
Polyethylene Terephthalates -
Stress, Mechanical -
Suture Techniques - instrumentation
Sutures -
Tensile Strength -
Wound Healing -

Find related publications in this database (Keywords)
Meniscus
Repair
Biomechanics
Load-to-failure
Stiffness
Devices