Rumpler, M; Suppan, M; Hajnsek, M; Satzinger, V; Kostler, S; Sinner, F.
Design and evaluation of a microfluidic-based whole blood pre-treatment module to detect folic acid
BIOMED ENG-BIOMED TE. 2013; 58: Doi: 10.1515/bmt-2013-4137
Web of Science FullText FullText_MUG

 

Führende Autor*innen der Med Uni Graz
Sinner Frank Michael
Altmetrics:

Dimensions Citations:
Plum Analytics:


Scite (citation analytics):

Abstract:
Haemoglobin interferes with the optical detection of various analytes in human blood. The aim of this study was the design and evaluation of a pre-treatment module which combines the instillation of undiluted whole human blood into a microchannel structure and the subsequent separation of erythrocytes. The microchannel structure on a cyclic olefin polymer (COP) based chip was manufactured by hot embossing or micro-milling. We were able to show that the used material does not adsorb haemoglobin or erythrocytes and that the generated plasma had 67.1% less haemoglobin than the whole blood sample.

Find related publications in this database (Keywords)
lab-on-a-chip
microfluidic
blood plasma separation
diagnostics
haemoglobin adsorption