Comparison of heart rate and oxygen saturation measurements from Masimo and Nellcor pulse oximeters in newly born term infants.
Acta Paediatr. 2013; 102(10):955-960 Doi: 10.1111/apa.12329
Web of Science PubMed FullText FullText_MUG

 

Co-authors Med Uni Graz
Schmölzer Georg
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
Aim To compare heart rate (HR) measurements from Masimo and Nellcor pulse oximeters (POs) against HR measured via a three lead electrocardiograph (ECG) (HRECG). We also compared peripheral oxygen saturation (SpO(2)) measurements between Nellcor and Masimo oximeters. Method Term infants born via elective caesarean section were studied. ECG leads were placed on the infant's chest and abdomen. Masimo and Nellcor PO sensors were randomly allocated to either foot. The monitors were placed on a trolley, and data from each monitor screen captured by a video camera. HR, SpO(2) measurements and signal quality were extracted. Bland-Altman analysis was used to determine agreement between HR from the ECG and each oximeter, and between SpO(2) from the oximeters. Results We studied 44 infants of whom 4 were resuscitated. More than 8000 pairs of observations were used for each comparison of HR and SpO(2.) The mean difference (+/- 2SD) between HRECG and HRNellcor was -0.8 (+/- 11) beats per minute (bpm); between HRECG and HRMasimo was 0.2 (+/- 9)bpm. The mean (+/- 2SD) difference between SpO(2Masimo) and SpO(2Nellcor) was -3 (+/- 15)%. The Nellcor PO measured 20% higher than the Masimo PO at SpO(2) <70%. Conclusion Both oximeters accurately measure HR. There was good agreement between SpO(2) measurements when SpO(2) 70%. At lower SpO(2), agreement was poorer.
Find related publications in this database (using NLM MeSH Indexing)
Biological Markers - blood
Cesarean Section -
Electrocardiography -
Female -
Heart Rate -
Humans -
Infant, Newborn -
Oximetry - instrumentation
Oxygen - blood
Pregnancy -
Surgical Procedures, Elective -
Term Birth -

Find related publications in this database (Keywords)
Delivery room
Electrocardiograph
Heart rate
Infant
Oxygen saturation
© Med Uni GrazImprint