Eisenhuber, E; Stadler, A; Prokop, M; Fuchsjager, M; Weber, M; Schaefer-Prokop, C.
Detection of monitoring materials on bedside chest radiographs with the most recent generation of storage phosphor plates: Dose increase does not improve detection performance
Radiology. 2003; 227(1):216-221 Doi: 10.1148/radiol.2271020045 [OPEN ACCESS]
Web of Science PubMed FullText FullText_MUG Google Scholar

 

Co-Autor*innen der Med Uni Graz
Fuchsjäger Michael
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
PURPOSE: To evaluate the performance of the most recent generation of storage phosphor plates for the detection of low-contrast catheter material on bedside chest radiographs. MATERIALS AND METHODS: In 10 patients in the intensive care unit, bedside chest radiographs were obtained with a 400-speed conventional screen-film system and with storage phosphor plates with exposure levels comparable to a 200-, 400-, or 800-speed conventional system. The chest radiograph was divided into 20 regions, 60% of which were superimposed with low-contrast catheter fragments. Six observers independently assessed the presence of catheter fragments by using a receiver operating characteristic (ROC) methodology. RESULTS: Detection performance (mean area under the ROC curve [Az]) with the storage phosphor plates was significantly superior to that with the screen-film system (Az = 0.76) at all three dose levels (Az = 0.88, 0.87, and 0.83 for 200-, 400-, and 800-speed doses, respectively; P <.05). Increasing the dose to a 200-speed system did not significantly increase detection performance compared with that with the 400-speed digital radiographs (Az = 0.88 vs 0.87). Dose reduction to 800 speed significantly deteriorated the detection performance (Az = 0.83) compared with that with the 400- and 200-speed digital radiographs, respectively. CONCLUSION: The most recent generation of storage phosphor plates is superior to a 400-speed screen-film system for the detection of catheter material, even at an exposure level of 800 speed.
Find related publications in this database (using NLM MeSH Indexing)
Adult -
Aged -
Aged, 80 and over -
Dose-Response Relationship, Radiation -
Equipment Design -
Female -
Humans -
Male -
Middle Aged -
Point-of-Care Systems -
Radiographic Image Enhancement -
Radiography, Thoracic - methods Radiography, Thoracic - standards
Radiology Information Systems -
X-Ray Intensifying Screens -

Find related publications in this database (Keywords)
radiography, bedside
radiography, comparative studies
radiography, storage phosphor
receiver operating characteristic (ROC) curve
thorax, radiography