Pabst, T; Sandstede, J; Beer, M; Kenn, W; Greiser, A; von Kienlin, M; Neubauer, S; Hahn, D.
Optimization of ECG-triggered 3D Na-23 MRI of the human heart
Magn Reson Med. 2001; 45(1):164-166 Doi: 10.1002/1522-2594(200101)45:1<164::AID-MRM1022>3.0.CO;2-S
Web of Science PubMed FullText FullText_MUG

 

Co-authors Med Uni Graz
Beer Meinrad
Altmetrics:

Dimensions Citations:

Plum Analytics:

Scite (citation analytics):

Abstract:
(23)Na MRI may allow distinction of normal and ischemically injured myocardium. The aim of this study was to optimize (23)Na MRI of the human heart by improvement of spatial resolution and ECG-triggering and to measure the signal/noise of blood and myocardium and the myocardium/blood signal ratios in a volunteer study. A spoiled gradient echo sequence was developed on a 1.5T scanner equipped with a (23)Na heart surface coil. 3D short axis ECG-triggered (23)Na MRI was performed in 10 healthy subjects. The signal/noise of myocardium and blood were 8.2 +/- 0.7 and 18.3 +/- 1.3, respectively, signal ratio myocardium/blood was 0.44 +/- 0.03. This value is in good agreement with the theoretical value of 0.45. ECG gated 3D (23)Na MRI of the human heart is feasible with sufficient spatial resolution and signal/noise ratio. Magn Reson Med 45:164-166, 2001.
Find related publications in this database (using NLM MeSH Indexing)
Adult -
Electrocardiography -
Female -
Heart - anatomy and histology
Humans -
Image Enhancement -
Magnetic Resonance Imaging - methods
Male -
Sodium -

Find related publications in this database (Keywords)
magnetic resonance imaging
Na-23 imaging
heart
signal/noise
© Med Uni GrazImprint